EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Acute Performance of the DynamX Bioadaptor in Complex Coronary Lesions: An Optical Coherence Tomography-based Evaluation

Arif Al Nooryani iD, Ahmed Khashaba iD, George Sianos iD, Nagwa Abdelrahman iD

Interventional Cardiology Reviews Research Resources · 28.04.2026

DOI: 10.15420/icr.2025.48 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Interventional Cardiology Reviews Research Resources, online erschienen am 28.04.2026. Verfasst von Arif Al Nooryani und 3 weiteren Autor:innen. Beteiligte Einrichtung: Cardiovascular Department, Al Qassimi Hospital, Emirates Health Services Sharjah, United Arab Emirates. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The DynamX Coronary Bioadaptor System incorporates a novel uncaging element that disengages approximately 6 months post-implantation following polymer resorption, while preserving longitudinal structural integrity. This innovative design aims to address the limitations of conventional drug-eluting stents by facilitating restoration of physiological vessel function in the treated segment without compromising radial support.

Methods

This retrospective study assessed the acute mechanical performance of the DynamX Bioadaptor in complex coronary lesions, as evaluated by optical coherence tomography (OCT) before and after DynamX implantation, with specific focus on lumen gain, device expansion, and strut apposition.

Results

A total of 29 coronary lesions from 24 patients were retrospectively analysed by OCT. Pre-implantation plaque morphology revealed ruptured plaques in 34.5% of lesions and calcified plaques in 55%. An OCT-based calcium score of 3 or 4 was observed in nine lesions. Post-implantation OCT demonstrated an acute lumen gain of 1.37 ± 0.8 mm and a residual area stenosis of 14.7 ± 9%. The incidence of malapposed struts (strut–lumen distance >300 µm) was low (0.64 ± 1.34%). Edge dissections were observed in four lesions, with two (6.9%) requiring bailout stenting.

Conclusion

The DynamX Bioadaptor showed favourable acute performance, achieving adequate lumen enlargement and optimal expansion and apposition in calcified and thrombotic lesion subsets, supporting its use in contemporary clinical practice.

Schlagwörter

Stichwörter der Autor:innen

Complex coronariesDynamX Bioadaptorexpansionlumen gainoptical coherence tomographystrut apposition

Autor:innen und Institutionen

  • Cardiovascular Department, Al Qassimi Hospital, Emirates Health Services Sharjah, United Arab Emirates.
  • Cardiovascular Department, Faculty of Medicine, Assiut University Assiut, Egypt.

ORCID-Profile: Arif Al Nooryani, Ahmed Khashaba, George Sianos, Nagwa Abdelrahman

Zitieren

APA

Nooryani, A. A., Khashaba, A., Sianos, G., & Abdelrahman, N. (2026). Acute Performance of the DynamX Bioadaptor in Complex Coronary Lesions: An Optical Coherence Tomography-based Evaluation. Interventional Cardiology Reviews Research Resources. https://doi.org/10.15420/icr.2025.48

Vancouver

Nooryani AA, Khashaba A, Sianos G, Abdelrahman N. Acute Performance of the DynamX Bioadaptor in Complex Coronary Lesions: An Optical Coherence Tomography-based Evaluation. Interventional Cardiology Reviews Research Resources. 2026. doi:10.15420/icr.2025.48

BibTeX
@article{Nooryani2026,
  author = {Nooryani, Arif Al and Khashaba, Ahmed and Sianos, George and Abdelrahman, Nagwa},