EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Preanalytical Variables Influence the Accuracy of Glucose Testing for Diagnostic Purposes

Oluwaseye Mary Onasanya, William A Anong

Integrative Journal of Medical Sciences · 23.02.2026 · 12

DOI: 10.15342/ijms.2026.752 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Integrative Journal of Medical Sciences, online erschienen am 23.02.2026. Verfasst von Oluwaseye Mary Onasanya und 1 weiteren Autor:innen. Die Arbeit nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The accuracy of test outcomes is essential for the quality of patient care. To demonstrate the effect of processing delays on glucose levels, we evaluate the stability and decline of glucose in whole blood samples obtained in EDTA vacutainers, which are stored at varying temperatures, as well as the rate of glucose degradation in different color-coded vacutainers.

Methods

Whole blood collected via venipuncture into the vacutainer was carefully mixed to ensure homogeneity. In the temperature study, glucose concentrations were measured at the outset and subsequently at 30-minute intervals in samples preserved at different temperatures. For the study involving color-coded vacutainers, glucose levels in the specimens were assessed initially and at various intervals.

Results

In general, the concentration of glucose in stored whole blood decreased over time. The decline was most significant in the specimen kept at 37 °C when compared to those stored at 24 °C and at ice (0 °C). In the color-coded analysis, all vacutainers, with the exception of the gray top, exhibited a statistically significant decrease in glucose levels. The gray top vacutainer retained as much as 94.7±2.0% of its original glucose concentration even after a duration of 24 hours.

Conclusion

The gray top vacutainer preserves glucose levels to ensure diagnostic precision in situations where a delay in specimen processing is anticipated.

Zitieren

APA

Onasanya, O. M., & Anong, W. A. (2026). Preanalytical Variables Influence the Accuracy of Glucose Testing for Diagnostic Purposes. Integrative Journal of Medical Sciences, 12. https://doi.org/10.15342/ijms.2026.752

Vancouver

Onasanya OM, Anong WA. Preanalytical Variables Influence the Accuracy of Glucose Testing for Diagnostic Purposes. Integrative Journal of Medical Sciences. 2026;12. doi:10.15342/ijms.2026.752

BibTeX
@article{Onasanya2026,
  author = {Onasanya, Oluwaseye Mary and Anong, William A},
  title = {Preanalytical Variables Influence the Accuracy of Glucose Testing for Diagnostic Purposes},
  journal = {Integrative Journal of Medical Sciences},
  year = {2026},
  volume = {12},
  doi = {10.15342/ijms.2026.752}
}