EUVIMEDEUVIMED entdecken
Crossref

Rubber Coring of Injectable Medication Vial Stoppers: An Evaluation of Causal Factors

Philip Chennell, Daniel Bourdeaux, Quentin Citerne, Catherine Arnaud, Sylvie Cosserant, Véra Boiko-Alaux, Valerie Sautou

Pharmaceutical Technology in Hospital Pharmacy · 2016

DOI: 10.1515/pthp-2016-0015 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Pharmaceutical Technology in Hospital Pharmacy, 2016. Verfasst von Philip Chennell und 6 weiteren Autor:innen. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 46 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Coring of a medication vial’s rubber stopper has been reported as a major cause of visible particle presence in injectable preparations. In this study, we investigated and quantified visible particle formation caused by coring associated with four potential causal factors.

Methods

The factors studied were: nature of the rubber stopper; rubber stopper thickness, type of metal needle bevel used to pierce the stopper, and puncture technique. For each one of 16 different situations, 40 medication vial rubber stoppers were punctured, and the contents filtered. The filters were then examined under optical microscopy and particles present counted and measured.

Results

The incidence of particle formation ranged from 0 % to 75 %, depending on the situation. Particle length was on average of 0.98±0.39 mm. The situation that gave the most particles (75 %; 30/40) was obtained when using a short bevelled needle, a 4 mm thick chlorobutyl vial stopper and with a puncture angle of 90°. Whilst a puncture technique reduced particle formation by more than 50 % for the most at risk situation, but without eliminating particle formation (residual formation of 22.5 %; 11/40), the use of a blunt bevelled needle totally eliminated the incidence of visible particle creation. The thickness of the rubber and the nature of the elastomer seemed to be linked to coring incidence, but in lesser proportions.

Conclusion

Puncturing the stoppers using a technique with a 45° puncture angle reduced particle formation, but only the use of a blunt metal needle totally eliminated it.

Zitieren

APA

Chennell, P., Bourdeaux, D., Citerne, Q., Arnaud, C., Cosserant, S., Boiko-Alaux, V., & Sautou, V. (2016). Rubber Coring of Injectable Medication Vial Stoppers: An Evaluation of Causal Factors. Pharmaceutical Technology in Hospital Pharmacy. https://doi.org/10.1515/pthp-2016-0015

Vancouver

Chennell P, Bourdeaux D, Citerne Q, Arnaud C, Cosserant S, Boiko-Alaux V, et al. Rubber Coring of Injectable Medication Vial Stoppers: An Evaluation of Causal Factors. Pharmaceutical Technology in Hospital Pharmacy. 2016. doi:10.1515/pthp-2016-0015

BibTeX
@article{Chennell2016,
  author = {Chennell, Philip and Bourdeaux, Daniel and Citerne, Quentin and Arnaud, Catherine and Cosserant, Sylvie and Boiko-Alaux, Véra and Sautou, Valerie},
  title = {Rubber Coring of Injectable Medication Vial Stoppers: An Evaluation of Causal Factors},
  journal = {Pharmaceutical Technology in Hospital Pharmacy},
  year = {2016},
  doi = {10.1515/pthp-2016-0015}
}
Kostenlos starten →