Fachartikel in der Zeitschrift Innovative Surgical Sciences, erschienen am 12.04.2017. Verfasst von Cas D P Van't Hullenaar und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgery, Meander Medical Centre, Maatweg 3, 3813 TZ Amersfoort, The Netherlands. Die Arbeit wurde bisher 31-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Robot-assisted surgery is considered to improve ergonomics over standard endoscopic surgery. Nevertheless, previous research demonstrated ergonomic deficits in the current console set-up.

Aim

This study was designed to objectively assess body posture in the da Vinci console during robot-assisted endoscopic surgery.

Methods

Multiple sagittal photographs from six physicians were taken during robot-assisted procedures. Trunk, neck, shoulder, elbow, hip, and knee angles were calculated and compared to ergonomic preferable joint angles. A 2D geometric model was developed using individual anthropometrics. Optimal seat height, armrest height, and viewer height were calculated. These results were compared to the findings of the sagittal photographs.

Results

Mean joint angles show potentially harmful neck angles for all participants. Trunk angles vary between surgeons, from inadequate to correct. In short and very tall individuals, optimal armrest height is outside the adjustment range of the console.

Conclusion

The da Vinci Surgical System console seating position results in a nonergonomic neck and trunk angle. The developed geometric model revealed that armrest height has a limited adjustment range. Adjustments to the console and optimization of preoperative settings are goals to further improve ergonomics in robot-assisted surgery.

Schlagwörter

Stichwörter der Autor:innen

ergonomicsoperating room environmentrobot-assisted surgery

Autor:innen und Institutionen

  • Department of Surgery, Meander Medical Centre, Maatweg 3, 3813 TZ Amersfoort, The Netherlands.
  • Department of Technical Medicine, Twente University, Noordhorst (NH101), Postbus 217, Enschede, The Netherlands.

Zitieren

APA

Van't Hullenaar, C. D. P., Hermans, B., & Broeders, I. A. M. J. (2017). Ergonomic assessment of the da Vinci console in robot-assisted surgery. Innovative Surgical Sciences, 2(2), 97-104. https://doi.org/10.1515/iss-2017-0007

Vancouver

Van't Hullenaar CDP, Hermans B, Broeders IAMJ. Ergonomic assessment of the da Vinci console in robot-assisted surgery. Innovative Surgical Sciences. 2017;2(2):97-104. doi:10.1515/iss-2017-0007

BibTeX
@article{VantHullenaar2017,
  author = {Van't Hullenaar, Cas D P and Hermans, Ben and Broeders, Ivo A M J},
  title = {Ergonomic assessment of the da Vinci console in robot-assisted surgery},
  journal = {Innovative Surgical Sciences},
  year = {2017},
  volume = {2},
  number = {2},
  pages = {97--104},
  doi = {10.1515/iss-2017-0007}
}
Kostenlos starten →