Incomplete filling of spray-dried K 2 EDTA evacuated blood tubes: impact on measuring routine hematological parameters on Sysmex XN-10
Diagnosis · 2023
DOI: 10.1515/dx-2023-0064 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnosis, 2023. Verfasst von Giuseppe Lippi und 6 weiteren Autor:innen. Beteiligte Einrichtung: Section of Clinical Biochemistry and School of Medicine, University of Verona, Verona, Italy. Hauptthemen laut MEDLINE: Anticoagulants, Hematology. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 18 Literaturangaben.
Abstract (Originalsprache)
Objectives
Because there is little published evidence on the effects of incomplete filling of K 2 EDTA evacuated blood tubes on routine hematological testing, this original study aimed to provide updated information on this preanalytical aspect.
Methods
The study population consisted of 17 ostensibly healthy volunteers. Blood was drawn by venipuncture with a 10 mL syringe and dispensed in varying amounts (0.2, 0.5, 1.0, 2.0, and 3.0 mL) into 3.0 mL blood tubes containing spray-dried 5.4 mg K 2 EDTA. All tubes were gently mixed and used to perform routine hematology tests on the Sysmex XN-10. Clinically significant variations were defined when the limits of desirable specifications of bias derived from biologic variation were exceeded.
Results
The desirable bias was exceeded in 33 % filled tubes (1.0 mL) for hematocrit and MCV (increased values) and for MCHC (decreased values), while it was exceeded in 17 % filled tubes (0.5 mL) for hemoglobin, hematocrit and MCV (increased values), and for MCHC (decreased values). Finally, the variation of values was higher than the desirable bias for RBC, hemoglobin, hematocrit and MCV (increase), and for MCHC and MPV (decrease) in 7 % filled tubes (0.2 mL). No clinically significant variations were observed in tubes filled up to 67 % of their nominal volume (i.e., 2.0 mL).
Conclusions
Consideration should be given to reject spray-dried K 2 EDTA blood tubes that contain a blood volume
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Section of Clinical Biochemistry and School of Medicine, University of Verona, Verona, Italy.
- Department of Haematology, Sydney Centres for Thrombosis and Haemostasis, Institute of Clinical Pathology and Medical Research (ICPMR), NSW Health Pathology, Westmead Hospital, Westmead, NSW Australia.
- School of Medical Sciences, Faculty of Medicine and Health, University of Sydney, Westmead Hospital, Westmead, NSW, Australia.
- Faculty of Science and Health, Charles Sturt University, Wagga Wagga, NSW, Australia.
ORCID-Profile: Giuseppe Lippi, Emmanuel J Favaloro
Zitieren
APA
Lippi, G., Dima, F., Tosi, M., Carpenè, G., Celegon, G., Favaloro, E. J., & Salvagno, G. L. (2023). Incomplete filling of spray-dried K 2 EDTA evacuated blood tubes: impact on measuring routine hematological parameters on Sysmex XN-10. Diagnosis. https://doi.org/10.1515/dx-2023-0064
Vancouver
Lippi G, Dima F, Tosi M, Carpenè G, Celegon G, Favaloro EJ, et al. Incomplete filling of spray-dried K 2 EDTA evacuated blood tubes: impact on measuring routine hematological parameters on Sysmex XN-10. Diagnosis. 2023. doi:10.1515/dx-2023-0064