Combined measurement uncertainty of HbA 1c and glucose using ISO 20914:2019 and outcome-based simulation

Emine Feyza Yurt iD, Medine Alpdemir iD, Mehmet Şeneş iD

Clinical Chemistry and Laboratory Medicine (CCLM) · 2026

DOI: 10.1515/cclm-2026-0388 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), 2026. Verfasst von Emine Feyza Yurt und 2 weiteren Autor:innen. Beteiligte Einrichtung: Departmant of Medical Biochemistry, Ankara Training and Research Hospital, Ankara, Türkiye. Hauptthemen laut MEDLINE: Glycated Hemoglobin, Blood Glucose. Die Arbeit nennt 23 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

This study aimed to calculate the measurement uncertainty (MU) of HbA 1c and glucose according to ISO/TS 20914:2019 guideline and to evaluate using indirect outcome-based analytical performance specifications (APS).

Methods

MU was estimated using internal quality control data collected between October 2023 and August 2025. A total of 13,270 and 13,824 IQC results for HbA 1c and 1,329 and 1,208 results for glucose were analyzed for Level 1 and Level 2 controls, respectively. HbA 1c was measured by capillary electrophoresis using two analyzers (12 capillaries each), and glucose was measured using the hexokinase method on three platforms. Maximum allowable uncertainty (MAU) targets were determined using the APS Simulator based on 2,000 patient samples with simultaneous HbA 1c and fasting plasma glucose results. Clinical decision limits were defined as 5.7 % (39 mmol/mol) and 6.5 % (48 mmol/mol) for HbA 1c , and 5.6 and 7 mmol/L for glucose. Overall agreement was used as the primary performance metric.

Results

Relative expanded MU (Urel) ranged from 4.56 to 4.59 % for HbA 1c expressed in %, 6.36–7.87 % for HbA 1c expressed in mmol/mol, and 3.68–4.36 % for glucose. Glucose Urel values remained below the minimum MAU limit (5.1 %), whereas HbA 1c values exceeded the corresponding limits (2.8 and 4.7 % for % and mmol/mol). Sensitivity and specificity ranged from 83.9 to 96 % and 93.4 to 99 % for glucose, and 85.4 to 93.4 % and 89 to 97.6 % for HbA 1c .

Conclusions

Glucose measurements met the minimum APS targets derived from the simulation model, whereas HbA 1c measurements exceeded these limits, indicating higher MU and a potential impact on patient classification around clinical decision limits.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Glycated HemoglobinBlood GlucoseHumansUncertaintyElectrophoresis, CapillaryQuality Control

Stichwörter der Autor:innen

HbA1cISO 20914:2019glucosemeasurement uncertaintyoutcome-based model

Substanzen

Glycated HemoglobinBlood Glucosehemoglobin A1c protein, human

Autor:innen und Institutionen

  • Departmant of Medical Biochemistry, Ankara Training and Research Hospital, Ankara, Türkiye.

ORCID-Profile: Emine Feyza Yurt, Medine Alpdemir, Mehmet Şeneş

Zitieren

APA

Feyza Yurt, E., Alpdemir, M., & Şeneş, M. (2026). Combined measurement uncertainty of HbA 1c and glucose using ISO 20914:2019 and outcome-based simulation. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2026-0388

Vancouver

Feyza Yurt E, Alpdemir M, Şeneş M. Combined measurement uncertainty of HbA 1c and glucose using ISO 20914:2019 and outcome-based simulation. Clinical Chemistry and Laboratory Medicine (CCLM). 2026. doi:10.1515/cclm-2026-0388

BibTeX
@article{FeyzaYurt2026,
  author = {Feyza Yurt, Emine and Alpdemir, Medine and Şeneş, Mehmet},
  title = {Combined measurement uncertainty of HbA 1c and glucose using ISO 20914:2019 and outcome-based simulation},
  journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
  year = {2026},
  doi = {10.1515/cclm-2026-0388}
}
Kostenlos starten →