Clinical validation of a DBS-based LC-MS/MS method for 25-hydroxyvitamin D: from lab sampling to home sampling
Clinical Chemistry and Laboratory Medicine (CCLM) · 2026
DOI: 10.1515/cclm-2025-1002 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), 2026. Verfasst von Liesl Heughebaert und 2 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Toxicology, Department of Bioanalysis, Faculty of Pharmaceutical Sciences, Ghent University, Ghent, Belgium. Hauptthemen laut MEDLINE: Tandem Mass Spectrometry, Dried Blood Spot Testing, Vitamin D. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 62 Literaturangaben.
Abstract (Originalsprache)
Objectives
As (a lack of) vitamin D has been linked to a wide variety of chronic diseases, there is growing interest in generating robust epidemiological data. Consequently, the need for large-scale biological sample collection has increased. In this context, dried blood spot (DBS) microsampling offers a minimally invasive alternative to venous sampling. However, to allow interpretation of DBS-based 25-hydroxyvitamin D (25-(OH)D) results, the set-up of a clinical validation study is essential to assess the agreement with the reference matrix, plasma.
Methods
Venous plasma and whole blood, venous DBS (vDBS) and capillary DBS (cDBS) were collected from 44 healthy volunteers to evaluate the agreement between the different sample types for 25-(OH)D quantification. To transform cDBS-based results to plasma concentrations, a hematocrit (Hct)-dependent conversion factor was applied and evaluated using four different Hct determination approaches. All samples were analysed using previously described validated LC-MS/MS methods.
Results
No clinically relevant methodological (vDBS vs. whole blood) or sampling-site related (cDBS vs. vDBS) issues were observed. Following Hct-dependent conversion, good agreement between the cDBS-derived and actually measured plasma results was obtained, as 90 % of the results lay within 20 % of the plasma result, independent of the Hct approach used. Additionally, weighted Cohen’s kappa values of 0.83–0.85 were obtained across the different Hct approaches, indicating substantial to almost perfect agreement in vitamin D status classification.
Conclusions
Following Hct-dependent conversion, cDBS can be used as a reliable and practical alternative matrix to plasma for large-scale monitoring of vitamin D status in epidemiological and public health contexts.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Laboratory of Toxicology, Department of Bioanalysis, Faculty of Pharmaceutical Sciences, Ghent University, Ghent, Belgium.
ORCID-Profile: Christophe P Stove
Zitieren
APA
Heughebaert, L., Ghesquière, R., & Stove, C. P. (2026). Clinical validation of a DBS-based LC-MS/MS method for 25-hydroxyvitamin D: from lab sampling to home sampling. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2025-1002
Vancouver
Heughebaert L, Ghesquière R, Stove CP. Clinical validation of a DBS-based LC-MS/MS method for 25-hydroxyvitamin D: from lab sampling to home sampling. Clinical Chemistry and Laboratory Medicine (CCLM). 2026. doi:10.1515/cclm-2025-1002
BibTeX
@article{Heughebaert2026,
author = {Heughebaert, Liesl and Ghesquière, Rosalie and Stove, Christophe P},
title = {Clinical validation of a DBS-based LC-MS/MS method for 25-hydroxyvitamin D: from lab sampling to home sampling},
journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
year = {2026},
doi = {10.1515/cclm-2025-1002}
}
EUVIMED