Commutability evaluation of glycated albumin candidate EQA materials
Clinical Chemistry and Laboratory Medicine (CCLM) · 2025
DOI: 10.1515/cclm-2025-0507 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), 2025. Verfasst von Rui Wu und 9 weiteren Autor:innen. Beteiligte Einrichtung: National Center for Clinical Laboratories, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of …. Hauptthemen laut MEDLINE: Serum Albumin. Die Arbeit nennt 19 Literaturangaben. Gefördert von: National Key Research and Development Program of China.
Abstract (Originalsprache)
Objectives
In clinical practice, glycated albumin (GA), as a key biomarker reflecting the level of blood glucose control in the short and medium term, has been widely used in the diagnosis and efficacy monitoring of diabetes. However, the consistency of GA results between different detection systems remains one of the current challenges in laboratory medicine. we evaluated the commutability of 10 evaluated samples to identify candidate EQA materials for GA measurement.
Methods
According to the Clinical and Laboratory Standards Institute (CLSI) document EP14-A3 and the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) commutability evaluation program, we used the established isotope dilution liquid chromatography-tandem mass spectrometry (ID LC-MS/MS) method for serum glycated albumin determination as the comparative method, and six different commercial kits used enzyme assay as the evaluating method. A total of 40 clinical sample serums, 9 pregnancy sample serums, and 10 evaluated samples were analyzed.
Results
For the CLSI approach, pooled serum samples (cRM1-3) at medium concentrations (cRM2) are commutable in 5/6 kits. For the IFCC approach, none of samples commutable.
Conclusions
The commutability evaluation results of the two approaches were different, and the cRM2 has commutability in most enzymatic kits. It can be a commutable material for the EQA project.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- National Center for Clinical Laboratories, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing, P.R. China.
- Chinese Academy of Medical Sciences, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, P.R. China.
- Department of Laboratory Medicine, Zhengzhou Central Hospital Affiliated to Zhengzhou University, Zhengzhou, P.R. China.
ORCID-Profile: Chuanbao Zhang
Förderung
- National Key Research and Development Program of China
Zitieren
APA
Wu, R., Zhou, H., Nizhamnuding, X., Pan, A., Zheng, H., Zhang, J., Zhao, H., Wang, J., Zhang, T., & Zhang, C. (2025). Commutability evaluation of glycated albumin candidate EQA materials. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2025-0507
Vancouver
Wu R, Zhou H, Nizhamnuding X, Pan A, Zheng H, Zhang J, et al. Commutability evaluation of glycated albumin candidate EQA materials. Clinical Chemistry and Laboratory Medicine (CCLM). 2025. doi:10.1515/cclm-2025-0507
BibTeX
@article{Wu2025,
author = {Wu, Rui and Zhou, Huijuan and Nizhamnuding, Xiaerbanu and Pan, Anqi and Zheng, Hao and Zhang, Jiangtao and Zhao, Haijian and Wang, Jing and Zhang, Tianjiao and Zhang, Chuanbao},
title = {Commutability evaluation of glycated albumin candidate EQA materials},
journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
year = {2025},
doi = {10.1515/cclm-2025-0507}
}
EUVIMED