Impact of delayed centrifugation on the stability of 32 biochemical analytes in blood samples collected in serum gel tubes and stored at room temperature
Clinical Chemistry and Laboratory Medicine (CCLM) · 2025
DOI: 10.1515/cclm-2025-0109 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), 2025. Verfasst von María Sanz-Felisi und 4 weiteren Autor:innen. Beteiligte Einrichtung: Territorial Clinical Laboratory Metropolitan South - Bellvitge University Hospital, Barcelona, Spain. Hauptthemen laut MEDLINE: Blood Specimen Collection, Temperature, Blood Chemical Analysis. Die Arbeit nennt 17 Literaturangaben.
Abstract (Originalsprache)
Objectives
Evaluate the stability of 32 biochemical analytes in venous blood samples stored at 18–25 °C under different time delays prior to centrifugation.
Methods
A prospective study was conducted involving 33 healthy volunteers. Four venous blood samples were collected from each participant. One sample was designated as baseline and processed immediately according to the tube manufacturer’s guidelines for centrifugation and analysis. The remaining three samples were stored under predefined conditions and centrifuged at different time intervals before undergoing analysis.
Results
Acceptable stability over the maximum storage time of 8 h was observed for 25 of the analytes tested in this study. However, direct bilirubin became unstable at 6 h and triglycerides at 8 h of storage prior to centrifugation. Calcium, gamma-glutamyl transferase, glucose, inorganic phosphate and potassium were found to be unstable in serum after 4 h of delayed centrifugation.
Conclusions
A delay in centrifugation of samples affected the stability of several analytes evaluated in the study, resulting in changes in their concentration or integrity. The analytical results for these analytes cannot be considered reliable as they do not meet the standards required for clinical validation. This underscores the importance of following stringent pre-analytical protocols to maintain the accuracy and reliability of laboratory diagnostic results.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Territorial Clinical Laboratory Metropolitan South - Bellvitge University Hospital, Barcelona, Spain.
Zitieren
APA
Sanz-Felisi, M., Tauler-Quetglas, P., Escartín-Díez, T., Arbiol-Roca, A., & Dot-Bach, D. (2025). Impact of delayed centrifugation on the stability of 32 biochemical analytes in blood samples collected in serum gel tubes and stored at room temperature. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2025-0109
Vancouver
Sanz-Felisi M, Tauler-Quetglas P, Escartín-Díez T, Arbiol-Roca A, Dot-Bach D. Impact of delayed centrifugation on the stability of 32 biochemical analytes in blood samples collected in serum gel tubes and stored at room temperature. Clinical Chemistry and Laboratory Medicine (CCLM). 2025. doi:10.1515/cclm-2025-0109
BibTeX
@article{SanzFelisi2025,
author = {Sanz-Felisi, María and Tauler-Quetglas, Paula and Escartín-Díez, Teresa and Arbiol-Roca, Ariadna and Dot-Bach, Dolors},
title = {Impact of delayed centrifugation on the stability of 32 biochemical analytes in blood samples collected in serum gel tubes and stored at room temperature},
journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
year = {2025},
doi = {10.1515/cclm-2025-0109}
}
EUVIMED