Verification, implementation and harmonization of automated chemiluminescent immunoassays for MPO- and PR3-ANCA detection
Clinical Chemistry and Laboratory Medicine (CCLM) · 2024
DOI: 10.1515/cclm-2023-0764 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), 2024. Verfasst von Manca Ogrič und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Rheumatology, University Medical Centre Ljubljana, Ljubljana, Slovenia. Hauptthemen laut MEDLINE: Antibodies, Antineutrophil Cytoplasmic, Anti-Neutrophil Cytoplasmic Antibody-Associated Vasculitis. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
Antineutrophil cytoplasmic antibody (ANCA) testing assists clinicians diagnose ANCA-associated vasculitis (AAV). We aimed to verify and harmonize chemiluminescent immunoassays for the detection of myeloperoxidase (MPO)- and proteinase 3 (PR3)-ANCA.
Methods
An in-house ELISA, a capture ELISA, and a chemiluminescent assay QUANTA Flash on a BIO-FLASH analyzer were used to detect MPO- and PR3-ANCA in sera from 39 patients with AAV, 55 patients with various non-AAV, and 66 patients with connective tissue diseases. The results of the assays were evaluated, and their clinical performance was assessed. The precision and linearity of the QUANTA Flash assays were determined, and likelihood ratios (LRs) for AAV at diagnosis were calculated.
Results
The precision and linearity of the QUANTA Flash assays were confirmed. Overall agreement between 97.5 and 98.8 % and Cohen’s kappa coefficients between 0.861 and 0.947 were observed for the results of the QUANTA Flash assays and ELISAs. The diagnostic sensitivity, specificity, and ROC analysis of the assays for AAV were statistically similar (in-house ELISA 89.7 %, 95.0 %, and 0.937; capture ELISA 92.3 %, 98.3 %, and 0.939; and QUANTA Flash 89.7 %, 95.9 %, and 0.972). For the QUANTA Flash assay results, the interval-specific LRs for AAV at diagnosis were: 0–8 CU had LR 0.08, 8–29 CU had LR 1.03, 29–121 CU had LR 7.76, 121–191 CU had LR 12.4, and >191 CU had LR ∞.
Conclusions
The QUANTA Flash MPO and PR3 assays provide precise and consistent results and have comparable clinical utility for AAV. The calculated LRs were consistent with published LRs, confirming the utility of LRs for harmonization of ANCA results.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Rheumatology, University Medical Centre Ljubljana, Ljubljana, Slovenia.
- University of Primorska, FAMNIT, Koper, Slovenia.
- Faculty of Medicine, University of Ljubljana, Ljubljana, Slovenia.
- Faculty of Pharmacy, University of Ljubljana, Ljubljana, Slovenia.
ORCID-Profile: Saša Čučnik
Zitieren
APA
Ogrič, M., Švec, T., Poljšak, K. M., Žigon, P., Hočevar, A., & Čučnik, S. (2024). Verification, implementation and harmonization of automated chemiluminescent immunoassays for MPO- and PR3-ANCA detection. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2023-0764
Vancouver
Ogrič M, Švec T, Poljšak KM, Žigon P, Hočevar A, Čučnik S. Verification, implementation and harmonization of automated chemiluminescent immunoassays for MPO- and PR3-ANCA detection. Clinical Chemistry and Laboratory Medicine (CCLM). 2024. doi:10.1515/cclm-2023-0764
BibTeX
@article{Ogri2024,
author = {Ogrič, Manca and Švec, Tinka and Poljšak, Katjuša Mrak and Žigon, Polona and Hočevar, Alojzija and Čučnik, Saša},
title = {Verification, implementation and harmonization of automated chemiluminescent immunoassays for MPO- and PR3-ANCA detection},
journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
year = {2024},
doi = {10.1515/cclm-2023-0764}
}
EUVIMED