State-based prognostics with similarity enhancement for CT X-ray tubes using operational logs
Biomedical Engineering / Biomedizinische Technik · 13.07.2026
DOI: 10.1515/bmt-2026-0184 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Engineering / Biomedizinische Technik, online erschienen am 13.07.2026. Verfasst von Changjun Chen und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Engineering, The Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, China. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
The tube is a core component of the CT system and is characterized by high cost and limited lifespan. This study aims to develop a method for estimating the remaining useful life (RUL) of CT X-ray tubes without requiring a preset failure threshold or prior knowledge of tube life.
Methods
To reduce the impact of varying scan conditions, different scan types are first analyzed, and the one with the most stable tube current distribution is selected. Then, a composite metric is applied to degradation parameters to obtain a candidate health indicator (HI). Next, the candidate HI is normalized using the training set and transformed into a percentage health indicator (PHI) to represent tube degradation. Finally, a state-based prognostics with similarity enhancement is applied to the PHI to estimate RUL.
Results
The proposed method is evaluated using real tube log data under leave-one-out cross-validation. The results demonstrate that it achieves optimal prediction performance and enables cross-tube RUL estimation. Moreover, the prediction results can reflect the actual degradation progress of the tube without requiring a preset failure threshold.
Conclusions
The proposed approach provides an effective and practical framework for CT X-ray tube RUL estimation, demonstrating good potential for practical medical equipment maintenance.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Clinical Engineering, The Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, China.
- College of Biomedical Engineering and Instrument Science, Zhejiang University, Hangzhou, China.
- College of Mechanical and Electrical Engineering, China Jiliang University, Hangzhou, China.
- Department of Clinical Engineering, Zhejiang Provincial People's Hospital, Hangzhou, China.
ORCID-Profile: Changjun Chen
Zitieren
APA
Chen, C., Wang, X., Qian, L., Wang, L., Ye, C., Zhu, F., Xu, P., Chu, Y., & Zou, Q. (2026). State-based prognostics with similarity enhancement for CT X-ray tubes using operational logs. Biomedical Engineering / Biomedizinische Technik. https://doi.org/10.1515/bmt-2026-0184
Vancouver
Chen C, Wang X, Qian L, Wang L, Ye C, Zhu F, et al. State-based prognostics with similarity enhancement for CT X-ray tubes using operational logs. Biomedical Engineering / Biomedizinische Technik. 2026. doi:10.1515/bmt-2026-0184
BibTeX
@article{Chen2026,
author = {Chen, Changjun and Wang, Xiangjie and Qian, Linze and Wang, Ling and Ye, Chenting and Zhu, Fengjie and Xu, Panpan and Chu, Yonghua and Zou, Quchao},
title = {State-based prognostics with similarity enhancement for CT X-ray tubes using operational logs},
journal = {Biomedical Engineering / Biomedizinische Technik},
year = {2026},
doi = {10.1515/bmt-2026-0184}
}