Cold-plasma for the decontamination of laparoscopic trocar incisions: In Vitro antimicrobial efficacy of a prototype applicator

Robert Bansemer iD, Veronika Hahn, Manfred Stieber, Stefan Horn, Thomas von Woedtke, Klaus-Dieter Weltmann

Biomedical Engineering / Biomedizinische Technik · 23.04.2026

DOI: 10.1515/bmt-2025-0416 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Engineering / Biomedizinische Technik, online erschienen am 23.04.2026. Verfasst von Robert Bansemer und 5 weiteren Autor:innen. Beteiligte Einrichtung: None Leibniz Institute for Plasma Science and Technology (INP) , Greifswald, Germany. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The antimicrobial efficacy and operating parameters of a cold plasma prototype were investigated. The device concept is intended for use on laparoscopic trocar incisions.

Methods

The device (50 mm length, 10 mm diameter, glass-ceramic body with two wire-wound electrodes) was operated with two gas compositions and in different environments. In vitro decontamination was tested on wet agar plates inoculated with Staphylococcus aureus (Gram-positive) and Escherichia coli (Gram-negative) as well as on inoculated stainless steel and polypropylene strips. A 5-min plasma exposure was applied.

Results

Both microorganisms were effectively inactivated on wet agar surfaces (inhibition zone assay: up to 36 mm) using different process gas compositions. Additionally, antimicrobial action was confirmed for S. aureus on stainless steel and polypropylene substrates. The prototype thus shows consistent decontamination performance across tested modes.

Conclusions

The plasma source offers a promising, minimally invasive adjunct for preventing surgical site infections during laparoscopic procedures. Further development, advanced biological models, and compliance with regulatory standards (e.g., DIN SPEC 91315) are required before clinical implementation.

Schlagwörter

Stichwörter der Autor:innen

dielectric barrier discharge (DBD)infectionplasma medicinesurgery

Autor:innen und Institutionen

  • None Leibniz Institute for Plasma Science and Technology (INP) , Greifswald, Germany.

ORCID-Profile: Robert Bansemer

Zitieren

APA

Bansemer, R., Hahn, V., Stieber, M., Horn, S., von Woedtke, T., & Weltmann, K. D. (2026). Cold-plasma for the decontamination of laparoscopic trocar incisions: In Vitro antimicrobial efficacy of a prototype applicator. Biomedical Engineering / Biomedizinische Technik. https://doi.org/10.1515/bmt-2025-0416

Vancouver

Bansemer R, Hahn V, Stieber M, Horn S, von Woedtke T, Weltmann KD. Cold-plasma for the decontamination of laparoscopic trocar incisions: In Vitro antimicrobial efficacy of a prototype applicator. Biomedical Engineering / Biomedizinische Technik. 2026. doi:10.1515/bmt-2025-0416

BibTeX
@article{Bansemer2026,
  author = {Bansemer, Robert and Hahn, Veronika and Stieber, Manfred and Horn, Stefan and von Woedtke, Thomas and Weltmann, Klaus-Dieter},
  title = {Cold-plasma for the decontamination of laparoscopic trocar incisions: In Vitro antimicrobial efficacy of a prototype applicator},
  journal = {Biomedical Engineering / Biomedizinische Technik},
  year = {2026},
  doi = {10.1515/bmt-2025-0416}
}
Kostenlos starten →