Verification of fCAL turbo Buhlmann reagent for determining serum calprotectin

Helena Čičak iD, Željka Šarčević, Ana Sruk iD, Frane Bukvić, Daria Pašalić iD, Lora Dukić iD, Ana-Maria Šimundić iD

Advances in Laboratory Medicine / Avances en Medicina de Laboratorio · 29.12.2025 · 6 (4) · 427-433

DOI: 10.1515/almed-2024-0201 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Advances in Laboratory Medicine / Avances en Medicina de Laboratorio, online erschienen am 30.06.2025. Verfasst von Helena Čičak und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Laboratory Diagnostics, University Hospital "Sveti Duh", Zagreb, Croatia. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The aim of this study was to verify Buhlmann fCAL turbo reagent, used for analysing calprotectin in stool, on serum samples.

Methods

Leftover patients’ serum samples were used. Verification protocol included: precision performed in accordance with Clinical and Laboratory Standards Institute (CLSI) EP15-A3 guidelines; comparison of Buhlmann fCAL turbo and serum GCAL Gentian reagents; verification of reference intervals (RI) proposed from literature (0.68–5.45 mg/L) in accordance with CLSI EP28-A3C guidelines; limit of blank (LoB) and limit of quantitation (LoQ) performed in accordance with CLSI EP17-A2 guidelines; linearity performed in accordance with CLSI EP6-A guidelines; carry-over and lastly, hook effect. Acceptance criteria were based on manufacturer specifications stated in the package insert for the fCAL turbo Buhlmann reagent designed for analysing calprotectin in stool samples.

Results

The CVs% for precision were within acceptance criteria, ≤10 %. Bland-Altman analysis showed the presence of a bias of 3.4 mg/L (95 % confidence interval 2.00–4.79). The equation for the Passing-Bablok regression was y= −0.01 (−0.08 to 0.06) + 0.37 (0.36–0.40)x, and it showed proportional difference. The obtained LoB was 0 mg/L, while the LoQ was 0.04 mg/L (CV% ≤ 20 %). The method was linear in the range 0.04–4.14 mg/L, and carry-over (0.1 %) and hook effect were not detected. The proposed RI in the literature was verified using 20 leftover patients’ samples.

Conclusions

Despite being declared for analysing stool samples only, Buhlmann fCAL turbo reagent can be used to determine serum calprotectin. Buhlmann fCAL turbo reagent is not comparable to the GCAL Gentian reagent.

Schlagwörter

Stichwörter der Autor:innen

calprotectinpreanalyticalserumverification

Autor:innen und Institutionen

  • Department of Medical Laboratory Diagnostics, University Hospital "Sveti Duh", Zagreb, Croatia.
  • Department of Neurology, University Hospital "Sveti Duh", Zagreb, Croatia.
  • Department of Orthopaedic Surgery and Traumatology, University Hospital "Sveti Duh", Zagreb, Croatia.
  • Department of Medical Chemistry, Biochemistry and Clinical Chemistry, University of Zagreb School of Medicine, Zagreb, Croatia.
  • Faculty of Pharmacy and Biochemistry, University of Zagreb, Zagreb, Croatia.

ORCID-Profile: Helena Čičak, Ana Sruk, Daria Pašalić, Lora Dukić, Ana-Maria Šimundić

Zitieren

APA

Čičak, H., Šarčević, Ž., Sruk, A., Bukvić, F., Pašalić, D., Dukić, L., & Šimundić, A. M. (2025). Verification of fCAL turbo Buhlmann reagent for determining serum calprotectin. Advances in Laboratory Medicine / Avances en Medicina de Laboratorio, 6(4), 427-433. https://doi.org/10.1515/almed-2024-0201

Vancouver

Čičak H, Šarčević Ž, Sruk A, Bukvić F, Pašalić D, Dukić L, et al. Verification of fCAL turbo Buhlmann reagent for determining serum calprotectin. Advances in Laboratory Medicine / Avances en Medicina de Laboratorio. 2025;6(4):427-433. doi:10.1515/almed-2024-0201

BibTeX
@article{iak2025,
  author = {Čičak, Helena and Šarčević, Željka and Sruk, Ana and Bukvić, Frane and Pašalić, Daria and Dukić, Lora and Šimundić, Ana-Maria},
  title = {Verification of fCAL turbo Buhlmann reagent for determining serum calprotectin},
  journal = {Advances in Laboratory Medicine / Avances en Medicina de Laboratorio},
  year = {2025},
  volume = {6},
  number = {4},
  pages = {427--433},
  doi = {10.1515/almed-2024-0201}
}
Kostenlos starten →