Fachartikel in der Zeitschrift Urogenital Tract Infection, erschienen am 31.08.2025. Verfasst von Young Kyu Han und 3 weiteren Autor:innen. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Sysmex Corporation.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate the diagnostic performance of the Sysmex UF-5000 flow cytometer in detecting acute bacterial prostatitis (ABP) compared to standard urine culture.Materials and Methods: This retrospective study analyzed 45 urine samples from patients with a clinical diagnosis of ABP. Each sample was evaluated using the UF-5000 to measure red blood cells (RBC), white blood cells (WBC), and bacterial counts, and the results were compared with those from standard urine culture and Gram staining. Receiver operating characteristic curves were generated, and sensitivity, specificity, positive predictive value (PPV), and negative predictive value were determined. Concordance between Gram classification by the UF-5000 and conventional Gram staining was also evaluated.

Results

Of the 45 patients, 84.4% had positive urine cultures. The bacterial count parameter demonstrated the highest diagnostic performance (area under the curve [AUC]=0.79; sensitivity, 89.5%; PPV, 91.9%), outperforming WBC (AUC=0.76) and RBC (AUC=0.55). The Gram classification flag showed an overall concordance of 85.7% with conventional Gram staining, with a concordance rate of 88% for Gram-negative organisms.

Conclusions

The Sysmex UF-5000 exhibited good concordance with urine culture for patients with ABP, particularly through the bacterial count parameter. Although it does not replace culture, the UF-5000 may serve as a rapid adjunctive tool to support early clinical decision-making in suspected ABP cases.

Förderung

  • Sysmex Corporation

Zitieren

APA

Han, Y. K., Lee, J. W., Park, H., & Choi, J. B. (2025). Diagnostic Utility of the Sysmex UF-5000 Flow Cytometer in Acute Bacterial Prostatitis: A Retrospective Pilot Study. Urogenital Tract Infection, 20(2), 107-113. https://doi.org/10.14777/uti.2550022011

Vancouver

Han YK, Lee JW, Park H, Choi JB. Diagnostic Utility of the Sysmex UF-5000 Flow Cytometer in Acute Bacterial Prostatitis: A Retrospective Pilot Study. Urogenital Tract Infection. 2025;20(2):107-113. doi:10.14777/uti.2550022011

BibTeX
@article{Han2025,
  author = {Han, Young Kyu and Lee, Jeong Woo and Park, Hae-Il and Choi, Jin Bong},
  title = {Diagnostic Utility of the Sysmex UF-5000 Flow Cytometer in Acute Bacterial Prostatitis: A Retrospective Pilot Study},
  journal = {Urogenital Tract Infection},
  year = {2025},
  volume = {20},
  number = {2},
  pages = {107--113},
  doi = {10.14777/uti.2550022011}
}
Kostenlos starten →