Hemodynamic optimization of a novel single-puncture dual-lumen coaxial catheter for arteriovenous fistula access: A computational fluid dynamics study

Cemal Kocaaslan

Northern Clinics of Istanbul · 2026

DOI: 10.14744/nci.2026.08522 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Northern Clinics of Istanbul, 2026. Verfasst von Cemal Kocaaslan. Beteiligte Einrichtung: Department of Cardiovascular Surgery, Istanbul Medeniyet University Faculty of Medicine, Istanbul, Turkiye. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE: Arteriovenous fistula (AVF) is the gold standard for hemodialysis access, but traditional dual-needle cannulation is limited by a restricted cannulation area and progressive stenosis; single-puncture systems may reduce these risks if hemodynamics can be optimized to prevent recirculation. This study aimed to design a novel single-puncture dual-lumen coaxial catheter and to evaluate its hemodynamic performance, flow efficiency, and recirculation rate using high-resolution computational fluid dynamics (CFD). METHODS: In this CFD and design-optimization study, a dual-lumen coaxial catheter with a unique fenestration geometry was designed and analyzed using SolidWorks Flow Simulation at a physiological temperature of 37°C. Boundary conditions included an AVF flow rate of 1000 mL/min and a dialysis pump speed of 300 mL/min. The primary endpoints were flow velocity, shear stress, and recirculation control. RESULTS: CFD simulation (1,030,641 cells) showed a maximum flow velocity of 12.69 m/s and a peak calculated shear stress of 2,602.45 Pa at the narrowest lumen section; because blood damage depends on both stress magnitude and exposure time, these computational data alone do not establish blood compatibility. Thermodilution analysis yielded a theoretical recirculation rate of 0.007% under the modeled conditions, and flow trajectories showed spatial separation of the venous jet from the arterial intake. CONCLUSION: Under the modeled steady-state conditions (AVF flow, 1000 mL/min; dialysis pump flow, 300 mL/min), CFD predicted stable separation of the return and aspiration streams and a theoretical recirculation rate of 0.007%. These proof-of-concept findings support further geometric optimization and physical testing; they do not establish clinical safety, efficacy, or superiority over standard dual-needle cannulation.

Schlagwörter

Stichwörter der Autor:innen

Arteriovenous fistulacomputational fluid dynamicsdialysishemodialysis catheterrecirculationsingle-puncture catheter

Autor:innen und Institutionen

  • Department of Cardiovascular Surgery, Istanbul Medeniyet University Faculty of Medicine, Istanbul, Turkiye.

Zitieren

APA

Kocaaslan, C. (2026). Hemodynamic optimization of a novel single-puncture dual-lumen coaxial catheter for arteriovenous fistula access: A computational fluid dynamics study. Northern Clinics of Istanbul. https://doi.org/10.14744/nci.2026.08522

Vancouver

Kocaaslan C. Hemodynamic optimization of a novel single-puncture dual-lumen coaxial catheter for arteriovenous fistula access: A computational fluid dynamics study. Northern Clinics of Istanbul. 2026. doi:10.14744/nci.2026.08522

BibTeX
@article{Kocaaslan2026,
  author = {Kocaaslan, Cemal},
  title = {Hemodynamic optimization of a novel single-puncture dual-lumen coaxial catheter for arteriovenous fistula access: A computational fluid dynamics study},
  journal = {Northern Clinics of Istanbul},
  year = {2026},
  doi = {10.14744/nci.2026.08522}
}
Kostenlos starten →