Uji Recovery Alat Rayto RT-1904C terhadap Pemeriksaan Glukosa Metode GOD-PAP

Alfino Validita Sidiq, Tiara Larashinta, Gervacia J Ratnawati, Ratih Indrawati, Slamet Slamet, Hendra B Sungkawa

Jurnal Pengelolaan Laboratorium Pendidikan · 25.09.2026 · 9 (1) · 1-8

DOI: 10.14710/jplp.9.1.1-8 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Jurnal Pengelolaan Laboratorium Pendidikan, online erschienen am 25.09.2026. Verfasst von Alfino Validita Sidiq und 5 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-SA 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The accuracy of blood glucose testing is crucial in clinical diagnostics. A recovery test is a key validation parameter for evaluating the performance and reliability of an analytical instrument. This study aimed to evaluate the performance of the Rayto RT-1904C analyzer for glucose testing using the GOD-PAP method via a recovery test.

Methods

This true experimental study used glucose standard samples spiked into normal and abnormal control serum at three concentrations (1%, 2%, 3%). Each concentration was measured with 9 replicates using the Rayto RT-1904C analyzer. The recovery percentage was calculated by comparing the measured value to the target value. Statistical analysis used the Mann-Whitney test.

Results

The mean recovery percentages for normal control serum at 1%, 2%, and 3% concentrations were 95.81%, 98.59%, and 95.47%, respectively. For abnormal control serum, the mean values were 103.02%, 101.49%, and 97.70%. All values were within the acceptable range of 95-105%. Although the Mann-Whitney test showed a significant difference (p

Zitieren

APA

Sidiq, A. V., Larashinta, T., Ratnawati, G. J., Indrawati, R., Slamet, S., & Sungkawa, H. B. (2026). Uji Recovery Alat Rayto RT-1904C terhadap Pemeriksaan Glukosa Metode GOD-PAP. Jurnal Pengelolaan Laboratorium Pendidikan, 9(1), 1-8. https://doi.org/10.14710/jplp.9.1.1-8

Vancouver

Sidiq AV, Larashinta T, Ratnawati GJ, Indrawati R, Slamet S, Sungkawa HB. Uji Recovery Alat Rayto RT-1904C terhadap Pemeriksaan Glukosa Metode GOD-PAP. Jurnal Pengelolaan Laboratorium Pendidikan. 2026;9(1):1-8. doi:10.14710/jplp.9.1.1-8

BibTeX
@article{Sidiq2026,
  author = {Sidiq, Alfino Validita and Larashinta, Tiara and Ratnawati, Gervacia J and Indrawati, Ratih and Slamet, Slamet and Sungkawa, Hendra B},
  title = {Uji Recovery Alat Rayto RT-1904C terhadap Pemeriksaan Glukosa Metode GOD-PAP},
  journal = {Jurnal Pengelolaan Laboratorium Pendidikan},
  year = {2026},
  volume = {9},
  number = {1},
  pages = {1--8},
  doi = {10.14710/jplp.9.1.1-8}
}
Kostenlos starten →