Characterization and Performance Evaluation of the TCRT5000 Reflective Optical Sensor for Early Detection of Flow Disturbances in an IoT-Based Peristaltic Infusion System

Hotromasari Dabukke, Flora Sijabat, Salomo Sijabat, Sri Ulina, Mhd Aldi Primasyukra

Journal of Physics and Its Applications · 31.08.2026 · 8 (3)

DOI: 10.14710/jpa.v8i3.31636 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Physics and Its Applications, online erschienen am 31.08.2026. Verfasst von Hotromasari Dabukke und 4 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-SA 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Flow disturbances in peristaltic infusion systems, including reduced drip rates, unstable dripping, and occlusion, may compromise infusion therapy and increase patient safety risks. This study aimed to characterize the performance of a TCRT5000 reflective optical sensor integrated into an Internet of Things (IoT)-based peristaltic infusion monitoring system for early detection of infusion flow disturbances. The proposed prototype consisted of a TCRT5000 reflective optical sensor for drip detection, a NodeMCU ESP32 for data acquisition and processing, Firebase for cloud-based data storage, and an MIT App Inventor mobile application for real-time monitoring. Experiments were conducted at infusion flow rates of 20, 50, 80, 100, 150, 200, 250, 300, 350, and 400 drops/min. Sensor performance was evaluated through sensitivity, linearity, and repeatability analyses. The results showed that the sensor exhibited a sensitivity ranging from 0.78 to 1.25 ADC/(drops/min) and a highly linear response described by the regression equation y = 1.0082x + 3.2590, with a coefficient of determination R² = 0.9992 and a correlation coefficient r = 0.9996. Repeatability analysis produced coefficients of variation below 5% for most measurements, indicating consistent sensor performance, although slightly higher variations were observed at lower flow rates. These findings demonstrate that the TCRT5000 reflective optical sensor provides reliable sensitivity, linearity, and repeatability, supporting its application as a sensing element in IoT-based infusion monitoring and the development of early infusion flow disturbance detection systems.

Zitieren

APA

Dabukke, H., Sijabat, F., Sijabat, S., Ulina, S., & Primasyukra, M. A. (2026). Characterization and Performance Evaluation of the TCRT5000 Reflective Optical Sensor for Early Detection of Flow Disturbances in an IoT-Based Peristaltic Infusion System. Journal of Physics and Its Applications, 8(3). https://doi.org/10.14710/jpa.v8i3.31636

Vancouver

Dabukke H, Sijabat F, Sijabat S, Ulina S, Primasyukra MA. Characterization and Performance Evaluation of the TCRT5000 Reflective Optical Sensor for Early Detection of Flow Disturbances in an IoT-Based Peristaltic Infusion System. Journal of Physics and Its Applications. 2026;8(3). doi:10.14710/jpa.v8i3.31636

BibTeX
@article{Dabukke2026,
  author = {Dabukke, Hotromasari and Sijabat, Flora and Sijabat, Salomo and Ulina, Sri and Primasyukra, Mhd Aldi},
  title = {Characterization and Performance Evaluation of the TCRT5000 Reflective Optical Sensor for Early Detection of Flow Disturbances in an IoT-Based Peristaltic Infusion System},
  journal = {Journal of Physics and Its Applications},
  year = {2026},
  volume = {8},
  number = {3},
  doi = {10.14710/jpa.v8i3.31636}
}
Kostenlos starten →