Automated analysis of spinal radiographs with calculation of sagittal balance parameters: development and validation of a clinical measurement support system
Russian Journal of Spine Surgery (Khirurgiya Pozvonochnika) · 08.10.2026 · 23 (3) · 68-75
DOI: 10.14531/ss2026.3.68-75 ↗Kurz gefasst
Fachartikel in der Zeitschrift Russian Journal of Spine Surgery (Khirurgiya Pozvonochnika), online erschienen am 08.10.2026. Verfasst von D. V. Ivanov und 9 weiteren Autor:innen. Die Arbeit nennt 19 Literaturangaben.
Abstract (Originalsprache)
Objective
To develop and evaluate a system for automated analysis of standing anteroposterior and lateral spinal radiographs with subsequent calculation of sagittal balance parameters and determination of personalized target spinopelvic parameters based on the pelvic incidence (PI). Material and Methods. A retrospective two-center study was conducted. The analysis included anonymized digital radiographs of the spine and pelvis in anteroposterior and lateral projections obtained in a standing position from 1,076 patients. A total of 973 fully annotated and expert-verified studies were used for training, validation, and testing. Automatic segmentation was performed using a U-Net architecture, followed by calculation of sagittal balance parameters with the possibility of deriving coronal parameters. A separate analysis of agreement between expert and automated measurements was conducted for the PI, PT, SS, GLL, and L4–S1 parameters. Segmentation quality was assessed using the Dice coefficient and IoU (Jaccard index). Inter-expert variability of annotations was additionally analyzed, and results were compared using “model–expert” and “expert–expert” schemes.
Results
On the test dataset, the mean Dice coefficient between the model and the expert was 90.1%, falling within the range of inter-expert variability (88.5–93.4%). A comparison of automated and expert measurements of PI, PT, SS, GLL, and L4–S1 parameters in a subset of 46 patients demonstrated high agreement between the methods: ICC(A,1) values ranged from 0.989 to 0.996; the 95% confidence intervals for ICC(A,1) ranged from 0.977 to 0.998; the mean absolute error ranged from 0.84° to 1.14°, and the proportion of measurements with an absolute difference of no more than 2.5° was 93.5–97.8%. The complete analysis cycle for a single patient required less than 3 seconds on a graphics processing unit (GPU). The prototype components successfully underwent pilot testing.
Conclusion
The developed system enables the automatic recognition of anatomical structures on spinal radiographs and the calculation of sagittal balance parameters with a high level of agreement with expert measurements. The system may be used to standardize radiographic measurements; however, its use in preoperative planning requires further clinical validation.
Zitieren
APA
Ivanov, D. V., Peretsmanas, E. O., Dol, A. V., Bessonov, L. V., Kovzalov, A. A., Rodin, A. A., Shchepeteva, Y. Y., Kuzminova, E. S., Smychkova, O. A., & Kisel, A. A. (2026). Automated analysis of spinal radiographs with calculation of sagittal balance parameters: development and validation of a clinical measurement support system. Russian Journal of Spine Surgery (Khirurgiya Pozvonochnika), 23(3), 68-75. https://doi.org/10.14531/ss2026.3.68-75
Vancouver
Ivanov DV, Peretsmanas EO, Dol AV, Bessonov LV, Kovzalov AA, Rodin AA, et al. Automated analysis of spinal radiographs with calculation of sagittal balance parameters: development and validation of a clinical measurement support system. Russian Journal of Spine Surgery (Khirurgiya Pozvonochnika). 2026;23(3):68-75. doi:10.14531/ss2026.3.68-75
BibTeX
@article{Ivanov2026,
author = {Ivanov, D. V. and Peretsmanas, E. O. and Dol, A. V. and Bessonov, L. V. and Kovzalov, A. A. and Rodin, A. A. and Shchepeteva, Yu. Yu. and Kuzminova, E. S. and Smychkova, O. A. and Kisel, A. A.},
title = {Automated analysis of spinal radiographs with calculation of sagittal balance parameters: development and validation of a clinical measurement support system},
journal = {Russian Journal of Spine Surgery (Khirurgiya Pozvonochnika)},
year = {2026},
volume = {23},
number = {3},
pages = {68--75},
doi = {10.14531/ss2026.3.68-75}
}
EUVIMED