Performance Evaluation of HemosIL Free Protein S

Eun Su Park, Hae In Bang, Jeong Won Shin, Rojin Park, Tae Yoon Choi

Clinical & Experimental Thrombosis and Hemostasis · 2015

DOI: 10.14345/ceth.15004 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical & Experimental Thrombosis and Hemostasis, 2015. Verfasst von Eun Su Park und 4 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

It is necessary to measure free protein S antigen levels in plasma for diagnosing protein S deficiency. The aim of this study was to evaluate the performance of HemosIL Free Protein S reagent (Instrumentation Laboratory, Lexington, MA, USA) which is an automated latex ligand immunoassay for the quantitative determination of free protein S.

Methods

HemosIL Free Protein S was evaluated for its precision, linearity, and reference range and compared to the HemosIL ProS, which is a reagent for the measurement of protein S activity in plasma, following the Clinical and Laboratory Standards Institute guidelines. The HemosIL Free Protein S was also compared to the STA-Liatest Free Protein S (Diagnostica Stago, Taverny, France), which is another reagent used in testing free protein S antigen levels.

Results

The within-run, between-run, and total coefficients of variation (CV) for HemosIL Free Protein S assay were 5.3%, 5.9%, and 7.9% at the low level, respectively, and 3.3%, 2.4%, and 4.1% at the high level, respectively. The linearity was in the range of 10.9-126.1% with a high correlation coefficient (R) of 0.997. The comparison between HemosIL Free protein S and ProS showed good correlation (R= 0.866). An excellent correlation (R= 0.931) was found between HemosIL Free protein S and STA-Liatest Free Protein S. The reference ranges for women and men were 50.8-116.9% and 69.4-138.3%, respectively.

Conclusion

HemosIL Free Protein S for measuring free protein S antigen in plasma had excellent precision and linearity. HemosIL Free Protein S was considered useful in detecting protein S deficiency. Keywords: Protein S deficiency, Free protein S, Precision, Linearity

Zitieren

APA

Park, E. S., Bang, H. I., Shin, J. W., Park, R., & Choi, T. Y. (2015). Performance Evaluation of HemosIL Free Protein S. Clinical & Experimental Thrombosis and Hemostasis. https://doi.org/10.14345/ceth.15004

Vancouver

Park ES, Bang HI, Shin JW, Park R, Choi TY. Performance Evaluation of HemosIL Free Protein S. Clinical & Experimental Thrombosis and Hemostasis. 2015. doi:10.14345/ceth.15004

BibTeX
@article{Park2015,
  author = {Park, Eun Su and Bang, Hae In and Shin, Jeong Won and Park, Rojin and Choi, Tae Yoon},
  title = {Performance Evaluation of HemosIL Free Protein S},
  journal = {Clinical & Experimental Thrombosis and Hemostasis},
  year = {2015},
  doi = {10.14345/ceth.15004}
}
Kostenlos starten →