The Lishtot TestDrop Pro device cannot reliably indicate microbial water safety

Timothy Purvis, Thao Nguyen, Caitlin McHugh, Siya Yeolekar, Eddy Ding, Juezihan Wang, Kevin Zhu, Joe Brown

PLOS Water · 18.06.2026 · 5 (6) · e0000442

DOI: 10.1371/journal.pwat.0000442 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS Water, online erschienen am 18.06.2026. Verfasst von Timothy Purvis und 7 weiteren Autor:innen. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Low-cost, scalable tests hold great potential to increase the availability of water safety data globally, especially in resource-limited settings. We evaluated the Lishtot TestDrop Pro, a commercially available, low-cost triboelectric-based sensor marketed for real-time determination of water contamination, including microbiological safety. We prepared a total of 199 water samples by serial dilution of raw wastewater influent into deionized water to produce a range of E. coli counts from 300 per dilution plate) CFU/100 mL to evaluate the sensor. We tested these waters using the manufacturer’s instructions. We collected sensor readings using three settings on each of five devices, generating nearly 8,955 individual measurements to compare against E. coli measured via standard membrane filtration assays (EPA Method 1604), a common metric of drinking water safety. We found no statistical correlation between the sensor score and E. coli presence in water: 43% of sensor readings indicated “safe” when wastewater and E. coli was present, and 56% of samples lacking E. coli were deemed “unsafe” by the sensor when wastewater was absent. The Lishtot TestDrop Pro is not an accurate method for measuring microbiological water safety.

Zitieren

APA

Purvis, T., Nguyen, T., McHugh, C., Yeolekar, S., Ding, E., Wang, J., Zhu, K., & Brown, J. (2026). The Lishtot TestDrop Pro device cannot reliably indicate microbial water safety. PLOS Water, 5(6), e0000442. https://doi.org/10.1371/journal.pwat.0000442

Vancouver

Purvis T, Nguyen T, McHugh C, Yeolekar S, Ding E, Wang J, et al. The Lishtot TestDrop Pro device cannot reliably indicate microbial water safety. PLOS Water. 2026;5(6):e0000442. doi:10.1371/journal.pwat.0000442

BibTeX
@article{Purvis2026,
  author = {Purvis, Timothy and Nguyen, Thao and McHugh, Caitlin and Yeolekar, Siya and Ding, Eddy and Wang, Juezihan and Zhu, Kevin and Brown, Joe},
  title = {The Lishtot TestDrop Pro device cannot reliably indicate microbial water safety},
  journal = {PLOS Water},
  year = {2026},
  volume = {5},
  number = {6},
  pages = {e0000442},
  doi = {10.1371/journal.pwat.0000442}
}
Kostenlos starten →