An operational key performance indicator framework and reference data from a multi-room pencil beam scanning proton therapy center
PLOS One · 2026
DOI: 10.1371/journal.pone.0360926 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift PLOS One, 2026. Verfasst von Jatinder Saini und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Washington School of Medicine, Seattle, Washington, United States of America. Hauptthemen laut MEDLINE: Proton Therapy, Quality Indicators, Health Care, Health Facilities, Health Facility Administration. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Multi-room proton therapy centers face unique operational challenges related to shared beam delivery systems, extended treatment hours, and high clinical complexity. This study presents a structured framework of key performance indicators (KPIs) and reports longitudinal single-institution operational reference data from a multi-room proton therapy center to support operational assessment and continuous quality improvement. Methods and materials Operational KPIs were retrospectively evaluated for calendar year 2024 at a three-room proton therapy center utilizing pencil beam scanning. KPIs were categorized into three areas: beam performance, treatment planning and pre-treatment workflow, and treatment room operations. Beam performance metrics included site- and room-specific uptime and beam pauses. Treatment planning metrics quantified CT simulation–to–patient scheduling timelines with attribution to physician, dosimetry, and physics components. Treatment room operations metrics included appointment variance, slot utilization, and waiting-for-beam (WFB) time. Data were collected automatically using a commercial analytics platform integrating radio-frequency identification tracking and oncology information system. Metrics were analyzed longitudinally and stratified by treatment room, disease site, and workflow component.
Results
For calendar year 2024, aggregate site uptime was 95.3%, weighted according to clinical operating time across the three treatment rooms; individual room uptimes ranged from 91.8% to 97.2%. Beam pauses were frequent but brief, resulting in 2,238 minutes of cumulative lost treatment time across 2024, equivalent to approximately 0.5% of total clinical operating time. Annual mean treatment planning time was 8.5 business days, with physician-related activities contributing the greatest variability across disease sites. Monthly average appointment variance ranged from 5 to 11 minutes, with most treatments initiated within 10 minutes of scheduled time. Across 2024, slot utilization was stable, with 73% of fractions completed within scheduled slots. Mean overrun and underrun durations were modest, although P90 overrun durations exceeded 10 minutes throughout the year. Mean WFB times were low and consistent, ranging from 1.5 to 2.3 minutes.
Conclusions
This study presents a practical KPI framework and a single-institution operational reference data for multi-room proton therapy center. These metrics may support longitudinal monitoring, identification of workflow inefficiencies, data-driven decision-making, and future multi-institutional benchmarking efforts.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Radiation Oncology, University of Washington School of Medicine, Seattle, Washington, United States of America.
- Fred Hutchinson Cancer Center-Proton Therapy, Seattle, Washington, United States of America.
ORCID-Profile: Jatinder Saini, Annemarie Shepherd
Zitieren
APA
Saini, J., Wong, T., Schwarz, M., McClure, R., Shepherd, A., Zeng, J., & Rengan, R. (2026). An operational key performance indicator framework and reference data from a multi-room pencil beam scanning proton therapy center. PLOS One. https://doi.org/10.1371/journal.pone.0360926
Vancouver
Saini J, Wong T, Schwarz M, McClure R, Shepherd A, Zeng J, et al. An operational key performance indicator framework and reference data from a multi-room pencil beam scanning proton therapy center. PLOS One. 2026. doi:10.1371/journal.pone.0360926
BibTeX
@article{Saini2026,
author = {Saini, Jatinder and Wong, Tony and Schwarz, Marco and McClure, Randall and Shepherd, Annemarie and Zeng, Jing and Rengan, Ramesh},
title = {An operational key performance indicator framework and reference data from a multi-room pencil beam scanning proton therapy center},
journal = {PLOS One},
year = {2026},
doi = {10.1371/journal.pone.0360926}
}
EUVIMED