Cable-routing geometry as a design variable for multiplanar hip assistance in a single-actuator soft exosuit
PLOS One · 29.10.2026 · 21 (10) · e0359731
DOI: 10.1371/journal.pone.0359731 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS One, online erschienen am 29.10.2026. Verfasst von Junhyeong Kwon und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Chung-Ang University, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Hip Joint, Hip. Die Arbeit nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation of Korea, Ministry of Trade, Industry and Energy.
Abstract (Originalsprache)
Cable-driven soft exosuits for gait assistance are predominantly sagittal-plane devices, yet pathological gait involves coordinated deviations across all three anatomical planes. This study experimentally evaluated whether cable-routing geometry alone, within a single-actuator hip exosuit, can produce measurable multiplanar hip effects during walking. Ten healthy, exosuit-experienced male adults walked on a treadmill under five conditions: one matched unpowered control and four powered configurations that paired single-plane and multi-plane cable routing with hip-flexion and hip-extension assistance. Three-dimensional hip kinematics were compared across conditions, and exosuit-generated hip torques were estimated from subject-specific musculoskeletal models. Multi-plane routing produced phase-dependent kinematic changes concentrated in the transverse plane. Relative to unassisted walking, multi-plane flexion assistance increased hip internal rotation (3.6°, p = 0.0002), multi-plane extension assistance increased hip external rotation (1.8°, p = 0.0095), and single-plane flexion assistance produced a smaller but significant increase in external rotation (1.1°, p = 0.0463). Frontal-plane differences were not significant for any comparison with unassisted walking, so the hypothesized larger frontal-plane deviations under multi-plane routing were not supported. Multi-plane routing also produced greater sagittal-plane hip excursion than single-plane routing. Overall gait-trajectory similarity between single-plane and multi-plane routing was comparable in most comparisons and differed significantly only in the hip frontal plane during extension assistance. The estimated torques indicated that multi-plane routing introduced larger frontal- and transverse-plane components while the sagittal-plane component remained dominant. These findings show that cable-routing geometry alone can shape phase-localized hip kinematics, reaching significance in the transverse plane, without adding actuators. The effect magnitudes were small and near the measurement uncertainty of the capture system. The all-male, exosuit-experienced sample limits generalizability, so relevance to clinical populations remains to be tested. Cable-routing geometry is therefore proposed as a candidate rather than an established design variable for multi-dimensional assistance in soft exosuits.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Mechanical Engineering, Chung-Ang University, Seoul, Republic of Korea.
- HUROTICS Inc., Seoul, Republic of Korea.
ORCID-Profile: Junhyeong Kwon, Giuk Lee
Förderung
- National Research Foundation of Korea
- Ministry of Trade, Industry and Energy
Zitieren
APA
Kwon, J., Park, S., Driscoll, H. G., & Lee, G. (2026). Cable-routing geometry as a design variable for multiplanar hip assistance in a single-actuator soft exosuit. PLOS One, 21(10), e0359731. https://doi.org/10.1371/journal.pone.0359731
Vancouver
Kwon J, Park S, Driscoll HG, Lee G. Cable-routing geometry as a design variable for multiplanar hip assistance in a single-actuator soft exosuit. PLOS One. 2026;21(10):e0359731. doi:10.1371/journal.pone.0359731
BibTeX
@article{Kwon2026,
author = {Kwon, Junhyeong and Park, Sejun and Driscoll, Harry George and Lee, Giuk},
title = {Cable-routing geometry as a design variable for multiplanar hip assistance in a single-actuator soft exosuit},
journal = {PLOS One},
year = {2026},
volume = {21},
number = {10},
pages = {e0359731},
doi = {10.1371/journal.pone.0359731}
}