Vivaray Hb pro non-invasive hemoglobin device: A prospective diagnostic accuracy study measuring agreement with a calibrated blood cell counter

Devesh Kumar, Sunita Kapoor, Ambanna Gowda iD, Dipti Gupta, Harsh Mittal, Suneet Sood iD

PLOS One · 03.11.2026 · 21 (11) · e0353640

DOI: 10.1371/journal.pone.0353640 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS One, online erschienen am 03.11.2026. Verfasst von Devesh Kumar und 5 weiteren Autor:innen. Beteiligte Einrichtung: IR Innovate Research Pvt. Ltd., Noida, Uttar Pradesh, India. Hauptthemen laut MEDLINE: Hemoglobins, Photoplethysmography, Hemoglobinometry. Die Arbeit nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Bosch Global Software Technologies.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Non-invasive hemoglobin measurement offers a painless and rapid alternative to conventional blood-based testing. The Vivaray Hb pro is a handheld photoplethysmography-based device designed for point-of-care hemoglobin assessment without blood sampling. We evaluated the clinical performance of the Vivaray Hb pro by comparing device-generated hemoglobin values with those obtained from a calibrated laboratory blood cell counter.

Methods

In this cross-sectional, non-randomized clinical performance study, participants aged ≥8 years were prospectively recruited. Hemoglobin was measured non-invasively using the Vivaray Hb pro and compared with venous blood samples analyzed on a calibrated Coulter counter. Agreement between methods was assessed using Bland–Altman analysis, including regression-based evaluation for proportional bias. Mean difference, mean absolute error (MAE) and proportions of measurements within tolerance limits were also calculated. Complete paired measurements were available for 763 individuals.

Results

Bland–Altman analysis returned a mean difference of −0.40 g/dL. The bias was hemoglobin-dependent, being higher than the true value at low hemoglobin levels and underestimation at higher levels. Regression-based analysis showed proportional bias (β 1 = −0.178), indicating decreasing difference with increasing hemoglobin concentration. The MAE was 1.5 g/dL but was lower (1.2) in the clinically predominant ranges of 8.1–13 g/dL.

Conclusion

The Vivaray Hb pro can function as a useful non-invsive screening and point-of-care assessment device for measuring hemoglobin, particularly in settings where rapid, painless and repeatable measurements are desirable.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HemoglobinsPhotoplethysmographyHemoglobinometryHumansFemaleProspective StudiesMaleMiddle AgedCross-Sectional StudiesAgedBlood Cell CountPoint-of-Care SystemsCalibrationAdult

Substanzen

Hemoglobins

Autor:innen und Institutionen

  • IR Innovate Research Pvt. Ltd., Noida, Uttar Pradesh, India.
  • City X-Ray and Scan Clinic Pvt. Ltd., New Delhi, India.
  • Citizen Hospital, Kalasipalya, Bengaluru, India.
  • Dr. Dipti's Multispeciality Clinics, Sahibabad, Uttar Pradesh, India.
  • Panchsheel Hospital, New Delhi, India.
  • Department of Surgery, IMU University, Malaysia.

ORCID-Profile: Ambanna Gowda, Suneet Sood

Förderung

  • Bosch Global Software Technologies

Zitieren

APA

Kumar, D., Kapoor, S., Gowda, A., Gupta, D., Mittal, H., & Sood, S. (2026). Vivaray Hb pro non-invasive hemoglobin device: A prospective diagnostic accuracy study measuring agreement with a calibrated blood cell counter. PLOS One, 21(11), e0353640. https://doi.org/10.1371/journal.pone.0353640

Vancouver

Kumar D, Kapoor S, Gowda A, Gupta D, Mittal H, Sood S. Vivaray Hb pro non-invasive hemoglobin device: A prospective diagnostic accuracy study measuring agreement with a calibrated blood cell counter. PLOS One. 2026;21(11):e0353640. doi:10.1371/journal.pone.0353640

BibTeX
@article{Kumar2026,
  author = {Kumar, Devesh and Kapoor, Sunita and Gowda, Ambanna and Gupta, Dipti and Mittal, Harsh and Sood, Suneet},
  title = {Vivaray Hb pro non-invasive hemoglobin device: A prospective diagnostic accuracy study measuring agreement with a calibrated blood cell counter},
  journal = {PLOS One},
  year = {2026},
  volume = {21},
  number = {11},
  pages = {e0353640},
  doi = {10.1371/journal.pone.0353640}
}
Kostenlos starten →