Evaluation of non-invasive hemoglobin measurement in blood donors at a tertiary care hospital, Bangkok, Thailand
PLOS One · 07.05.2026 · 21 (5) · e0348447
DOI: 10.1371/journal.pone.0348447 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS One, online erschienen am 07.05.2026. Verfasst von Ratinan Dangwilailert und 4 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand. Hauptthemen laut MEDLINE: Blood Donors, Hemoglobins, Hemoglobinometry. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Faculty of Medicine Siriraj Hospital, Mahidol University.
Abstract (Originalsprache)
Background
Hemoglobin (Hb) assessment is crucial to prevent blood donation from anemic donors. Most measurements are invasive and painful; however, an alternative, non-invasive hemoglobin measurement is available. This study aimed to compare non-invasive and point-of-care invasive hemoglobin test results with those from an automated blood analyzer in blood donors.
Methods
Qualified blood donors were enrolled at a tertiary care hospital in Bangkok, Thailand. Hemoglobin was initially measured by the non-invasive device (Rad-67 Pulse CO-Oximeter) and a point-of-care invasive hemoglobin screening device (Mission HemoPro). Participants with point-of-care invasive hemoglobin ≥ 12.5 g/dL were eligible to donate, and during donation, standard hemoglobin measurements were obtained using an automated analyzer, XN-550. Those who were ineligible with point-of-care invasive hemoglobin < 12.5 g/dL underwent additional blood sampling for hemoglobin measurement by an automated analyzer. Correlation and agreement between non-invasive, point-of-care, and automated hemoglobin measurements were assessed with Intraclass Correlation Coefficient (ICC) and Bland-Altman plots, while satisfaction was evaluated using a Likert scale.
Results
Of 300 participants, 295 had complete data. Of these, 169 were male (57.28%). Average non-invasive, point-of-care, and automated hemoglobin levels were 14.38 ± 1.12 g/dL, 13.65 ± 0.70 g/dL, and 13.90 ± 1.16 g/dL, respectively. The ICC between non-invasive, point-of-care, and automated hemoglobin measurement was 0.600 (95% CI: 0.522–0.668) and 0.897 (95% CI: 0.872–0.957). The sensitivity of the non-invasive Hb measurement was poor for detecting anemic donors when the Hb cut-off was set at 12.5 g/dL. This method received significantly higher satisfaction than the routinely used invasive device.
Conclusion
The non-invasive Hb measurement in blood donors showed moderate agreement with the standard test, but the sensitivity was poor when the cut-off hemoglobin was set at 12.5 g/dl. Since donor satisfaction was higher, this method may be used as an alternative screening tool, provided a higher Hb cut-off value is used.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
- Department of Transfusion Medicine, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
- Research Department, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
ORCID-Profile: Parichart Permpikul
Förderung
- Faculty of Medicine Siriraj Hospital, Mahidol University
Zitieren
APA
Dangwilailert, R., Lekmak, S., Promlee, D., Pongkunakorn, T., & Permpikul, P. (2026). Evaluation of non-invasive hemoglobin measurement in blood donors at a tertiary care hospital, Bangkok, Thailand. PLOS One, 21(5), e0348447. https://doi.org/10.1371/journal.pone.0348447
Vancouver
Dangwilailert R, Lekmak S, Promlee D, Pongkunakorn T, Permpikul P. Evaluation of non-invasive hemoglobin measurement in blood donors at a tertiary care hospital, Bangkok, Thailand. PLOS One. 2026;21(5):e0348447. doi:10.1371/journal.pone.0348447
BibTeX
@article{Dangwilailert2026,
author = {Dangwilailert, Ratinan and Lekmak, Somboon and Promlee, Duangtida and Pongkunakorn, Tanyaporn and Permpikul, Parichart},
title = {Evaluation of non-invasive hemoglobin measurement in blood donors at a tertiary care hospital, Bangkok, Thailand},
journal = {PLOS One},
year = {2026},
volume = {21},
number = {5},
pages = {e0348447},
doi = {10.1371/journal.pone.0348447}
}