Biomechanical optimization of iliolumbar fixation strategies for unilateral vertical sacral fractures: Prioritizing stability-mobility balance via finite element analysis
PLOS One · 2026
DOI: 10.1371/journal.pone.0339705 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS One, 2026. Verfasst von Yupeng Ma und 6 weiteren Autor:innen. Beteiligte Einrichtung: The First Ward of Trauma Orthopedics, Yantai Shan Hospital, Yantai, Shandong Province, People's Republic of China. Hauptthemen laut MEDLINE: Sacrum, Lumbar Vertebrae, Fracture Fixation, Internal, Spinal Fractures. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Science & Technology Innovation Development Project of Yantai City, China, National Natural Science Foundation of China, Shandong Provincial Key R&D Program of China.
Themen: Mobilisation
Abstract (Originalsprache)
Objective
This study aims to optimize iliolumbar fixation strategies for unilateral vertical sacral fractures via finite element analysis, by comparing stability, implant stress, and lumbar mobility to identify the optimal clinical option.
Methods
A finite element model of the pelvis and L3-L5 lumbar spine was constructed to simulate four fixation models (L4L5IS, L5S1IS, L5IS, S1IS) under a 600 N vertical load. Sacral vertical displacement, implant stress, and fracture line separation were analyzed.
Results
Double-segment fixation (L4 + L5 + iliac screw and L5 + S1 + iliac screw) demonstrated superior sacral stability compared to single-segment fixation (L5 + iliac screw and S1 + iliac screw). The L5 + S1 + iliac screw configuration achieved the best balance of stability and lumbar mobility. Stress concentrations were primarily observed at iliac screw connectors, but all models remained within safe mechanical limits.
Conclusions
Double-segment fixation, particularly the L5 + S1 + iliac screw model, is recommended for optimal sacral stability. For cases with compromised S1 pedicles, L4 + L5 + iliac screw fixation is a reliable alternative. Short-segment fixation is viable when prioritizing lumbar mobility.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- The First Ward of Trauma Orthopedics, Yantai Shan Hospital, Yantai, Shandong Province, People's Republic of China.
- Yantai Key Laboratory for Repair and Reconstruction of Bone & Joint, Yantai Shan Hospital, Yantai, Shandong Province, People's Republic of China.
- Orthopaedics Department, Yantai Shan Hospital, Yantai, Shandong Province, People's Republic of China.
- Yantai Zhifu District Center for Disease Control and Prevention, Shandong Yantai, Shandong Province, People's Republic of China.
ORCID-Profile: Yu Li
Förderung
- Science & Technology Innovation Development Project of Yantai City, China
- National Natural Science Foundation of China
- Shandong Provincial Key R&D Program of China
- Medical and Health Technology Development Program of Shandong Province, China
- Distinguished Middle-Aged and Young Scientist Encourage and Reward Foundation of Shandong Province
- Binzhou Medical University “Clinical+X” Scientific and Technological Innovation Project
- Binzhou Medical University “Clinical+X”
Zitieren
APA
Ma, Y., Liu, W., Huang, T., Hong, H., Zhao, Y., Xu, G., & Li, Y. (2026). Biomechanical optimization of iliolumbar fixation strategies for unilateral vertical sacral fractures: Prioritizing stability-mobility balance via finite element analysis. PLOS One. https://doi.org/10.1371/journal.pone.0339705
Vancouver
Ma Y, Liu W, Huang T, Hong H, Zhao Y, Xu G, et al. Biomechanical optimization of iliolumbar fixation strategies for unilateral vertical sacral fractures: Prioritizing stability-mobility balance via finite element analysis. PLOS One. 2026. doi:10.1371/journal.pone.0339705
BibTeX
@article{Ma2026,
author = {Ma, Yupeng and Liu, Weiwei and Huang, Tao and Hong, Huanyu and Zhao, Yong and Xu, Guofeng and Li, Yu},
title = {Biomechanical optimization of iliolumbar fixation strategies for unilateral vertical sacral fractures: Prioritizing stability-mobility balance via finite element analysis},
journal = {PLOS One},
year = {2026},
doi = {10.1371/journal.pone.0339705}
}
EUVIMED