EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Active expiration reduces hypercapnia in lung failure – results of the prospective interventional ActiveEx study and development of a prototype device for automated application

Denis Witham, Julius Valentin Kunz iD, Ann-Christin Krebs iD, Karoline Brückner, Roland Körner, Mareen Pigorsch iD, Kai-Uwe Eckardt, Paul Uwe Thamsen, Philipp Enghard, Julija Peter

PLOS One · 2025

DOI: 10.1371/journal.pone.0333579 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS One, 2025. Verfasst von Denis Witham und 9 weiteren Autor:innen. Beteiligte Einrichtung: Charité - Universitätsmedizin Berlin, corporate member of Freie Universität Berlin, Humboldt Universität zu Berlin, and Berlin Institute of Health; D…. Hauptthemen laut MEDLINE: Hypercapnia, Respiratory Insufficiency, Exhalation, Respiration, Artificial. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

This study investigates the efficacy of Intermittent Abdominal Pressure Ventilation (IAPV) and Expiratory Rib Cage Compression (ERCC) in reducing hypercapnia among critically ill, mechanically ventilated patients. It also assesses the feasibility of automating these techniques using a self-developed, ventilator-synchronized device on a dummy.

Methods

This single-arm feasibility study was conducted in two intensive care units at Charité – Universitätsmedizin Berlin, including critically ill patients with hypercapnic lung failure, with additional lab testing at the Technical University of Berlin. Manual IAPV and ERCC were applied to patients, and automation feasibility was tested on a dummy using prototype device. Primary outcomes included changes in tidal volume and partial pressure of carbon dioxide (PaCO2) and device effectiveness at different PEEP levels.

Results

In nine hypercapnic patients, manual IAPV increased tidal volume from 5.72 to 8.85 mL/ kg predicted bodyweight (p < 0.001, relative effect (CI) 0.93 (0.79–1.06)) and ERCC from 5.79 to 8.13 mL/ kg predicted bodyweight (p < 0.001, relative effect (CI) 0.93 (0.80–1.05)). PaCO2 reduced after 20 minutes with both techniques (IAPV: from 65 to 52 mmHg, p < 0.01, relative effect (CI) 0.15 (0.01–0.28); ERCC: from 61 to 51 mmHg, p= < 0.01, relative effect (CI) 0.22 (0.07–0.37)). A transient decrease in oxygenation was fully and rapidly reversible. The automated device doubled tidal volumes in dummy simulations, with greater effectiveness at higher PEEP-levels.

Conclusion

Manual and automated ventilator-synchronized IAPV and ERCC were associated with improved ventilation. Their potential role in managing hypercapnic respiratory failure—such as in weaning failure, obstructive lung disease, or neuromuscular weakness—remains a subject for future clinical research. German Registry of Clinical Trials (DRKS00027397)

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HypercapniaRespiratory InsufficiencyExhalationRespiration, ArtificialHumansMaleFemaleAgedMiddle AgedProspective StudiesTidal VolumeCarbon DioxideFeasibility StudiesPositive-Pressure RespirationIntensive Care UnitsCritical Illness

Substanzen

Carbon Dioxide

Autor:innen und Institutionen

  • Charité - Universitätsmedizin Berlin, corporate member of Freie Universität Berlin, Humboldt Universität zu Berlin, and Berlin Institute of Health; Department of Nephrology and Medical Intensive Care, 10117 Berlin, Germ…
  • Technical University of Berlin - Institute of Fluid Mechanics and Technical Acoustics of the Technical University of Berlin, Berlin, Germany.
  • Charité - Universitätsmedizin Berlin, corporate member of Freie Universität Berlin, Humboldt Universität zu Berlin, Institute of Biometry and Clinical Epidemiology, 10117 Berlin, Germany.

ORCID-Profile: Julius Valentin Kunz, Ann-Christin Krebs, Mareen Pigorsch

Zitieren

APA

Witham, D., Kunz, J. V., Krebs, A. C., Brückner, K., Körner, R., Pigorsch, M., Eckardt, K. U., Thamsen, P. U., Enghard, P., & Peter, J. (2025). Active expiration reduces hypercapnia in lung failure – results of the prospective interventional ActiveEx study and development of a prototype device for automated application. PLOS One. https://doi.org/10.1371/journal.pone.0333579

Vancouver

Witham D, Kunz JV, Krebs AC, Brückner K, Körner R, Pigorsch M, et al. Active expiration reduces hypercapnia in lung failure – results of the prospective interventional ActiveEx study and development of a prototype device for automated application. PLOS One. 2025. doi:10.1371/journal.pone.0333579

BibTeX
@article{Witham2025,
  author = {Witham, Denis and Kunz, Julius Valentin and Krebs, Ann-Christin and Brückner, Karoline and Körner, Roland and Pigorsch, Mareen and Eckardt, Kai-Uwe and Thamsen, Paul Uwe and Enghard, Philipp and Peter, Julija},
  title = {Active expiration reduces hypercapnia in lung failure – results of the prospective interventional ActiveEx study and development of a prototype device for automated application},
  journal = {PLOS One},
  year = {2025},
  doi = {10.1371/journal.pone.0333579}
}
Kostenlos starten →