EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Laser-based pulse oximetry eliminates pigmentation effects on oxygen saturation measurements: A pilot study

Jonas A Pologe iD, Nathan K You iD, Mark Blumstein, Kristine L Snyder, William W Hay

PLOS One · 06.10.2025 · 20 (10) · e0333109

DOI: 10.1371/journal.pone.0333109 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS One, online erschienen am 06.10.2025. Verfasst von Jonas A Pologe und 4 weiteren Autor:innen. Beteiligte Einrichtung: Research and Development, Zynex Monitoring Solutions, Boulder, Colorado, United States of America. Hauptthemen laut MEDLINE: Oximetry, Oxygen Saturation, Lasers, Pigmentation. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Zynex Monitoring Solutions.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Almost since its introduction, pulse oximetry has been plagued by inaccuracy associated with pigmentation, whether from fingernail polishes or melanin. The presence of melanin in the optical path of a pulse oximetry sensor has been shown to artifactually increase oxygen saturation measurements which can result, clinically, in occult hypoxemia and misdiagnoses.

Methods

This report describes the theoretical basis for this inaccuracy and presents results from both a benchtop study and a clinical study testing the effects of pigmentation on conventional LED-based pulse oximetry compared to laser-based pulse oximetry. The clinical portion of this study was performed on 18 consenting participants, nine darkly pigmented and nine lightly pigmented, to assess the ability of laser-based pulse oximetry to eliminate this dangerous pigmentation bias. The clinical study directly compared oxygen saturation readings on laser-based pulse oximeters to readings performed on two different LED-based pulse oximeters. All measurements were compared to invasive reference laboratory measurements performed on arterial blood samples. We hypothesized that monochromatic light sources used in laser-based pulse oximetry would make this new technology insensitive to pigmentation bias.

Results

The clinical portion of this study showed significantly greater (p < 0.001) measurement error (variance) for the two LED-based pulse oximeters (5.48 and 5.47) compared to laser-based pulse oximetry (3.54), when analyzed for all participants. The bias differences in oxygen saturation measurements by the LED-based pulse oximeters, when made on lightly versus darkly pigmented participants compared to invasive reference measurements, accounts for most of the increase in measurement error.

Conclusions

By combining theoretical development, experimental benchtop testing, and a clinical study, this research explains and demonstrates that the wide spectral bandwidth of LEDs is the root cause of pigmentation bias in commercially available LED-based pulse oximetry and validates the ability of narrow-band laser-based pulse oximetry to eliminate this pigmentation bias.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

OximetryOxygen SaturationLasersPigmentationOxygenHumansPilot ProjectsMaleFemaleAdultMiddle AgedMelaninsSkin Pigmentation

Substanzen

OxygenMelanins

Autor:innen und Institutionen

  • Research and Development, Zynex Monitoring Solutions, Boulder, Colorado, United States of America.
  • Denver, Colorado, United States of America.

ORCID-Profile: Jonas A Pologe, Nathan K You

Förderung

  • Zynex Monitoring Solutions

Zitieren

APA

Pologe, J. A., You, N. K., Blumstein, M., Snyder, K. L., & Hay, W. W. (2025). Laser-based pulse oximetry eliminates pigmentation effects on oxygen saturation measurements: A pilot study. PLOS One, 20(10), e0333109. https://doi.org/10.1371/journal.pone.0333109

Vancouver

Pologe JA, You NK, Blumstein M, Snyder KL, Hay WW. Laser-based pulse oximetry eliminates pigmentation effects on oxygen saturation measurements: A pilot study. PLOS One. 2025;20(10):e0333109. doi:10.1371/journal.pone.0333109

BibTeX
@article{Pologe2025,
  author = {Pologe, Jonas A and You, Nathan K and Blumstein, Mark and Snyder, Kristine L and Hay, William W},
  title = {Laser-based pulse oximetry eliminates pigmentation effects on oxygen saturation measurements: A pilot study},
  journal = {PLOS One},
  year = {2025},
  volume = {20},
  number = {10},
  pages = {e0333109},
  doi = {10.1371/journal.pone.0333109}
}
Kostenlos starten →