The efficacy of Netrod Six-electrode Radiofrequency Renal Denervation system: mathematical modeling, phantom experiments and animal studies
PLOS One · 2025
DOI: 10.1371/journal.pone.0323336 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS One, 2025. Verfasst von Xiaohua Song und 3 weiteren Autor:innen. Beteiligte Einrichtung: College of Chemistry and Life Science, Beijing University of Technology, Beijing, China. Hauptthemen laut MEDLINE: Kidney, Models, Theoretical, Denervation, Catheter Ablation. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China.
Abstract (Originalsprache)
The principle of Netrod TM Renal Denervation RF Generator is based on the heating of radiofrequency electrodes to achieve the therapeutic temperature for peripheral nerve ablation with minimal impact on intra-arterial blood flow. The extravascular heat distribution and continuous ablation were evaluated using COMSOL Multiphysics with an idealized 3D model and validated in an ex vivo experiment. The safety of ablation was verified by H&E staining of a hypertensive dog model in the treatment group of animal experiments, and changes in blood pressure before and after the procedure(assessed at 7, 30, and 60 days in animal experiments) were used to characterize the efficacy. The simulation results showed that the maximum injury depth range of the Netrod TM Six-electrode is 5.08mm and the depth of the ablation zone obtained in the Phantom experiment was 3–6mm. Circumferential coverage was up to 91.99% and the ablation zone was continuous. In animal experiments, eight animals in the treatment group had significantly lower blood pressure (88.9%). Nerve cells in nerve bundles showed vacuolar changes and thickening of peripheral sheaths after peripheral nerve ablation of renal arteries.The Netrod TM Six-electrode presented here has potential to effectively and safely ablate renal nerves. The demonstrated efficacy and safety in animal models provide a foundation for clinical trials in humans.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- College of Chemistry and Life Science, Beijing University of Technology, Beijing, China.
- Industrial Informatization and Science and Technology Bureau of Changqing District, Jinan, China.
- Shanghai Golden Leaf MedTech CO., Ltd, Shanghai, China.
ORCID-Profile: Qun Nan
Förderung
- National Natural Science Foundation of China
Zitieren
APA
Song, X., Guo, J., Cao, H., & Nan, Q. (2025). The efficacy of Netrod Six-electrode Radiofrequency Renal Denervation system: mathematical modeling, phantom experiments and animal studies. PLOS One. https://doi.org/10.1371/journal.pone.0323336
Vancouver
Song X, Guo J, Cao H, Nan Q. The efficacy of Netrod Six-electrode Radiofrequency Renal Denervation system: mathematical modeling, phantom experiments and animal studies. PLOS One. 2025. doi:10.1371/journal.pone.0323336
BibTeX
@article{Song2025,
author = {Song, Xiaohua and Guo, Jiulin and Cao, Hongguang and Nan, Qun},
title = {The efficacy of Netrod Six-electrode Radiofrequency Renal Denervation system: mathematical modeling, phantom experiments and animal studies},
journal = {PLOS One},
year = {2025},
doi = {10.1371/journal.pone.0323336}
}