Vergleichsstudie in der Zeitschrift PLOS ONE, 2025. Verfasst von Kainan Shao und 7 weiteren Autor:innen. Beteiligte Einrichtung: Cancer Center, Department of Radiation Oncology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital), Hangzhou Medical College, Hang…. Hauptthemen laut MEDLINE: Lung Neoplasms, Radiotherapy, Intensity-Modulated, Radiotherapy Planning, Computer-Assisted, Carcinoma, Non-Small-Cell Lung. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Science Foundation of Zhejiang Province, Medical Science and Technology Project of Zhejiang Province, Basic Public Welfare Research Program of Zhejiang Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

The dosimetric characteristics and treatment efficiency of VMAT plans using two linear accelerator platforms, Halcyon and Infinity, in conventional radiotherapy for non-small cell lung cancer (NSCLC) are compared to provide data for selecting clinical equipment. The study also explores potential confounding factors that may influence treatment outcomes.

Methods

This retrospective cohort study aims to compare the dosimetric characteristics and treatment efficiency of VMAT plans delivered using Halcyon and Infinity linear accelerator platforms in patients with NSCLC. A retrospective analysis was performed on 60 NSCLC patients receiving conventional fractionated radiotherapy with VMAT plans developed for both Halcyon and Infinity. These plans were optimized with RayStation 9A with identical dose constraints and optimization parameters. The groups were compared in terms of target dose coverage, normal tissue sparing, plan complexity, and treatment efficiency. The dosimetric parameters included D98%, D2%, and Dmean for both the CTV and PTV and dose distributions for organs at risk (OARs), including the heart, lungs, and spinal cord. Logistic regression was performed to account for potential confounding factors, such as PTV volume, tumor stage, and tumor location.

Results

The VMAT plans of both platforms met the clinical dosimetric requirements. Halcyon showed superior protection of normal tissues in low-dose areas (e.g., Lungs V5Gy and Heart V30Gy), whereas Infinity excelled in controlling hot spots and achieving rapid dose fall-off at the target margins. Furthermore, Halcyon has fewer plan monitoring units and lower complexity than Infinity and reduced treatment time by 24.0%. Logistic regression analysis revealed that PTV volume was a significant predictor for dose metric differences, while tumor stage and tumor location had variable effects depending on the dose metric, highlighting the need to account for these factors in clinical comparisons. Overall, there was no significant difference in target dose coverage or uniformity between the platforms; each demonstrated specific strengths in protecting different OARs and in treatment execution efficiency.

Conclusion

Halcyon and Infinity offer distinct advantages in radiotherapy for NSCLC. Halcyon provides better protection of normal tissues and performance in low-dose regions, whereas Infinity offers greater treatment efficiency and superior control in high-dose regions. The study also highlights that PTV volume is an important factor influencing dosimetric outcomes. In choosing optimal radiotherapy equipment in clinical practice, the study results suggest that treatment planning should leverage the unique technical features of different accelerators to achieve the best individualized outcomes. Future studies should increase the sample size and employ prospective research designs to confirm the clinical relevance of these findings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lung NeoplasmsRadiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-AssistedCarcinoma, Non-Small-Cell LungHumansRetrospective StudiesFemaleMaleMiddle AgedAgedRadiotherapy DosageOrgans at RiskParticle Accelerators

Autor:innen und Institutionen

  • Cancer Center, Department of Radiation Oncology, Zhejiang Provincial People's Hospital (Affiliated People's Hospital), Hangzhou Medical College, HangZhou, Zhejiang, China.
  • Department of Radiation Physics, Zhejiang Cancer Hospital, HangZhou, Zhejiang, China.

ORCID-Profile: Kainan Shao, Lingyun Qiu

Förderung

  • Natural Science Foundation of Zhejiang Province
  • Medical Science and Technology Project of Zhejiang Province
  • Basic Public Welfare Research Program of Zhejiang Province

Zitieren

APA

Shao, K., Du, F., Qiu, L., Zhang, Y., Li, Y., Ding, J., Zhan, W., & Chen, W. (2025). Comparative analysis of VMAT plans on Halcyon and infinity for lung cancer radiotherapy. PLOS ONE. https://doi.org/10.1371/journal.pone.0318462

Vancouver

Shao K, Du F, Qiu L, Zhang Y, Li Y, Ding J, et al. Comparative analysis of VMAT plans on Halcyon and infinity for lung cancer radiotherapy. PLOS ONE. 2025. doi:10.1371/journal.pone.0318462

BibTeX
@article{Shao2025,
  author = {Shao, Kainan and Du, Fenglei and Qiu, Lingyun and Zhang, Yinghao and Li, Yucheng and Ding, Jieni and Zhan, Wenming and Chen, Weijun},
  title = {Comparative analysis of VMAT plans on Halcyon and infinity for lung cancer radiotherapy},
  journal = {PLOS ONE},
  year = {2025},
  doi = {10.1371/journal.pone.0318462}
}
Kostenlos starten →