Fachartikel in der Zeitschrift PLOS ONE, 2024. Verfasst von Asma Alahmadi und 3 weiteren Autor:innen. Beteiligte Einrichtung: King Abdullah Ear Specialist Center (KAESC), King Saud Medical City, King Saud University, Riyadh, Saudi Arabia. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Software, Cochlear Implantation. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Anatomy-based fitting (ABF) for cochlear implant users is a new era that seeks improved outcomes. Recently, different imaging modalities, such as plain X-rays, have been proposed to build the ABF as an alternative to the computed tomography (CT) scan. This study aimed to assess the feasibility and validity of OTOPLAN ® software in building ABF using plain X-ray imaging. Patients and methods A retrospective evaluation of postoperative CT scans and plain X-ray post-op images of 54 patients was analyzed using the OTOPLAN ® software. The post-op analysis was done for the angular insertion depth (AID) and center frequency of each electrode contact using both imaging modalities. Moreover, inter-rater reliability was assessed for measurements obtained from CT scans and plain X-ray images.

Results

Non-significant statistical and clinical mismatches were detected when comparing the AID and center frequency measurements assessed using CT and X-rays. The absolute difference between CT and X-ray approaches ranged from 0.0 to 4.6 degrees for AID and 0.2 to 0.5 semitone for frequency. Moreover, the AID and the frequency measurements from CT and X-ray images demonstrated almost perfect agreement between the raters. The inter-observer reliability for CT scans showed that the intraclass correlation coefficient (ICC) exceeded 0.97 for AID and 0.95 for the frequency across all electrode contacts.

Conclusion

Our results demonstrated the validity and reliability of using post-operative X-ray images by OTOPLAN ® software to build Anatomy-based Fitting maps.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedSoftwareCochlear ImplantationHumansRetrospective StudiesCochlear ImplantsFemaleMaleReproducibility of ResultsMiddle AgedAdultAged

Autor:innen und Institutionen

  • King Abdullah Ear Specialist Center (KAESC), King Saud Medical City, King Saud University, Riyadh, Saudi Arabia.
  • Research Department, MED-EL GmbH, Riyadh, Saudi Arabia.

ORCID-Profile: Asma Alahmadi

Zitieren

APA

Alahmadi, A., Abdelsamad, Y., Hafez, A., & Hagr, A. (2024). X-ray guided anatomy-based fitting: The validity of OTOPLAN. PLOS ONE. https://doi.org/10.1371/journal.pone.0313567

Vancouver

Alahmadi A, Abdelsamad Y, Hafez A, Hagr A. X-ray guided anatomy-based fitting: The validity of OTOPLAN. PLOS ONE. 2024. doi:10.1371/journal.pone.0313567

BibTeX
@article{Alahmadi2024,
  author = {Alahmadi, Asma and Abdelsamad, Yassin and Hafez, Ahmed and Hagr, Abdulrahman},
  title = {X-ray guided anatomy-based fitting: The validity of OTOPLAN},
  journal = {PLOS ONE},
  year = {2024},
  doi = {10.1371/journal.pone.0313567}
}
Kostenlos starten →