EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Advanced low-power filter architecture for biomedical signals with adaptive tuning

Ramasamy Srinivasagan iD

PLOS ONE · 23.01.2025 · 20 (1) · e0311768

DOI: 10.1371/journal.pone.0311768 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS ONE, online erschienen am 23.01.2025. Verfasst von Ramasamy Srinivasagan. Beteiligte Einrichtung: Computer Engineering, CCSIT, King Faisal University, Al Hufuf, Kingdom of Saudi Arabia. Hauptthemen laut MEDLINE: Signal Processing, Computer-Assisted, Electrocardiography, Electroencephalography. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: King Faisal University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This paper presents a low-power, second-order composite source-follower-based filter architecture optimized for biomedical signal processing, particularly ECG and EEG applications. Source-follower-based filters are recommended in the literature for high-frequency applications due to their lower power consumption when compared to filters with alternative topologies. However, they are not suitable for biomedical applications requiring low cutoff frequencies as they are designed to operate in the saturation region. The major contribution in this work are the filter is made to operate in the weak inversion zone to reduce the area needed for the capacitor and the amount of power dissipated. Process variation is one of the major issues in the weak inversion regime. To overcome this, a unique method of compensating against fluctuations in process, voltage, and temperature is put forth based on magnitude comparison is another contribution. Key findings from post-layout simulations and experimental measurements demonstrate that the filter achieves a tunable cutoff frequency range of 0.5 Hz to 150 Hz, with a total power dissipation of only 6nW at 150 Hz. The design occupies a compact silicon area of 0.065 mm 2 and offers a dynamic range of 75 dB. The measured results indicate that for a 300 mVpp signal swing, the top bound on THD is -40 dB. The filter’s robustness against process, voltage, and temperature variations is validated through on-chip tuning using a current steering DAC, ensuring stable performance across different operating conditions. These results make the proposed filter a promising candidate for low-power biomedical devices. The recommended filter is developed and implemented using UMC-0.18μm CMOS technology with a 1.0V supply, and the IC is tapped out using an MPW run of Euro practice IC services.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Signal Processing, Computer-AssistedElectrocardiographyElectroencephalographyHumansAlgorithmsEquipment Design

Autor:innen und Institutionen

  • Computer Engineering, CCSIT, King Faisal University, Al Hufuf, Kingdom of Saudi Arabia.

ORCID-Profile: Ramasamy Srinivasagan

Förderung

  • King Faisal University

Zitieren

APA

Srinivasagan, R. (2025). Advanced low-power filter architecture for biomedical signals with adaptive tuning. PLOS ONE, 20(1), e0311768. https://doi.org/10.1371/journal.pone.0311768

Vancouver

Srinivasagan R. Advanced low-power filter architecture for biomedical signals with adaptive tuning. PLOS ONE. 2025;20(1):e0311768. doi:10.1371/journal.pone.0311768

BibTeX
@article{Srinivasagan2025,
  author = {Srinivasagan, Ramasamy},
  title = {Advanced low-power filter architecture for biomedical signals with adaptive tuning},
  journal = {PLOS ONE},
  year = {2025},
  volume = {20},
  number = {1},
  pages = {e0311768},
  doi = {10.1371/journal.pone.0311768}
}