Chassis-based fiber-coupled optical probe design for reproducible quantitative diffuse optical spectroscopy measurements

Giselle C Matlis, Qihuang Zhang iD, Emilie J Benson, M Katie Weeks, Kristen Andersen iD, Jharna Jahnavi, Alec Lafontant, Jake Breimann, Thomas Hallowell iD, Yuxi Lin, Daniel J Licht, Arjun G Yodh und 6 weitere

PLOS ONE · 2024

DOI: 10.1371/journal.pone.0305254 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS ONE, 2024. Verfasst von Giselle C Matlis und 17 weiteren Autor:innen. Beteiligte Einrichtung: Division of Neurology, Department of Pediatrics, Children's Hospital of Philadelphia, Philadelphia, PA, United States of America. Hauptthemen laut MEDLINE: Optical Fibers, Cerebrovascular Circulation. Die Arbeit nennt 81 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NINDS NIH HHS, NIBIB NIH HHS, NICHD NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Advanced optical neuromonitoring of cerebral hemodynamics with hybrid diffuse optical spectroscopy (DOS) and diffuse correlation spectroscopy (DCS) methods holds promise for non-invasive characterization of brain health in critically ill patients. However, the methods’ fiber-coupled patient interfaces (probes) are challenging to apply in emergent clinical scenarios that require rapid and reproducible attachment to the head. To address this challenge, we developed a novel chassis-based optical probe design for DOS/DCS measurements and validated its measurement accuracy and reproducibility against conventional, manually held measurements of cerebral hemodynamics in pediatric swine (n = 20). The chassis-based probe design comprises a detachable fiber housing which snaps into a 3D-printed, circumferential chassis piece that is secured to the skin. To validate its reproducibility, eight measurement repetitions of cerebral tissue blood flow index (BFI), oxygen saturation (StO 2 ), and oxy-, deoxy- and total hemoglobin concentration were acquired at the same demarcated measurement location for each pig. The probe was detached after each measurement. Of the eight measurements, four were acquired by placing the probe into a secured chassis, and four were visually aligned and manually held. We compared the absolute value and intra-subject coefficient of variation (CV) of chassis versus manual measurements. No significant differences were observed in either absolute value or CV between chassis and manual measurements (p > 0.05). However, the CV for BFI (mean ± SD: manual, 19.5% ± 9.6; chassis, 19.0% ± 10.8) was significantly higher than StO 2 (manual, 5.8% ± 6.7; chassis, 6.6% ± 7.1) regardless of measurement methodology (p

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Optical FibersCerebrovascular CirculationAnimalsSwineReproducibility of ResultsSpectrum AnalysisBrainEquipment DesignHemodynamicsHemoglobinsOxygen

Substanzen

HemoglobinsOxygen

Autor:innen und Institutionen

  • Division of Neurology, Department of Pediatrics, Children's Hospital of Philadelphia, Philadelphia, PA, United States of America.
  • School of Biomedical Engineering, Science and Health Systems, Drexel University, Philadelphia, Pennsylvania, United States of America.
  • Department of Biostatistics and Epidemiology, University of Pennsylvania, Philadelphia, PA, United States of America.
  • Department of Epidemiology, Biostatistics and Occupational Health, McGill University, Montreal, QC, Canada.
  • Department of Physics and Astronomy, University of Pennsylvania, Philadelphia, PA, United States of America.
  • Department of Anesthesiology and Critical Care Medicine, Children's Hospital of Philadelphia, Philadelphia, PA, United States of America.

ORCID-Profile: Qihuang Zhang, Kristen Andersen, Thomas Hallowell, Rodrigo M Forti, Wesley B Baker, Tiffany S Ko

Förderung

  • NINDS NIH HHS
  • NIBIB NIH HHS
  • NICHD NIH HHS
  • NHLBI NIH HHS
  • National Institutes of Health
  • Children’s Hospital of Philadelphia Frontier Program
  • American Heart Association

Zitieren

APA

Matlis, G. C., Zhang, Q., Benson, E. J., Weeks, M. K., Andersen, K., Jahnavi, J., Lafontant, A., Breimann, J., Hallowell, T., Lin, Y., Licht, D. J., Yodh, A. G., Kilbaugh, T. J., Forti, R. M., White, B. R., Baker, W. B., Xiao, R., & Ko, T. S. (2024). Chassis-based fiber-coupled optical probe design for reproducible quantitative diffuse optical spectroscopy measurements. PLOS ONE. https://doi.org/10.1371/journal.pone.0305254

Vancouver

Matlis GC, Zhang Q, Benson EJ, Weeks MK, Andersen K, Jahnavi J, et al. Chassis-based fiber-coupled optical probe design for reproducible quantitative diffuse optical spectroscopy measurements. PLOS ONE. 2024. doi:10.1371/journal.pone.0305254

BibTeX
@article{Matlis2024,
  author = {Matlis, Giselle C and Zhang, Qihuang and Benson, Emilie J and Weeks, M Katie and Andersen, Kristen and Jahnavi, Jharna and Lafontant, Alec and Breimann, Jake and Hallowell, Thomas and Lin, Yuxi and Licht, Daniel J and Yodh, Arjun G and Kilbaugh, Todd J and Forti, Rodrigo M and White, Brian R and Baker, Wesley B and Xiao, Rui and Ko, Tiffany S},
  title = {Chassis-based fiber-coupled optical probe design for reproducible quantitative diffuse optical spectroscopy measurements},
  journal = {PLOS ONE},
  year = {2024},
  doi = {10.1371/journal.pone.0305254}
}
Kostenlos starten →