EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A remote-controlled automatic chest compression device capable of moving compression position during CPR: A pilot study in a mannequin and a swine model of cardiac arrest

Gil Joon Suh iD, Taegyun Kim, Kyung Su Kim iD, Woon Yong Kwon iD, Hayoung Kim, Heesu Park, Gaonsorae Wang, Jaeheung Park, Sungmoon Hur, Jaehoon Sim iD, Kyunghwan Kim, Jung Chan Lee iD und 5 weitere

PLOS ONE · 2024

DOI: 10.1371/journal.pone.0297057 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS ONE, 2024. Verfasst von Gil Joon Suh und 16 weiteren Autor:innen. Beteiligte Einrichtung: Department of Emergency Medicine, Seoul National University College of Medicine, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Cardiopulmonary Resuscitation, Heart Arrest. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministry of Health and Welfare.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Recently, we developed a chest compression device that can move the chest compression position without interruption during CPR and be remotely controlled to minimize rescuer exposure to infectious diseases. The purpose of this study was to compare its performance with conventional mechanical CPR device in a mannequin and a swine model of cardiac arrest. Materials and methods A prototype of a remote-controlled automatic chest compression device (ROSCER) that can change the chest compression position without interruption during CPR was developed, and its performance was compared with LUCAS 3 in a mannequin and a swine model of cardiac arrest. In a swine model of cardiac arrest, 16 male pigs were randomly assigned into the two groups, ROSCER CPR (n = 8) and LUCAS 3 CPR (n = 8), respectively. During 5 minutes of CPR, hemodynamic parameters including aortic pressure, right atrial pressure, coronary perfusion pressure, common carotid blood flow, and end-tidal carbon dioxide partial pressure were measured.

Results

In the compression performance test using a mannequin, compression depth, compression time, decompression time, and plateau time were almost equal between ROSCER and LUCAS 3. In a swine model of cardiac arrest, coronary perfusion pressure showed no difference between the two groups ( p = 0.409). Systolic aortic pressure and carotid blood flow were higher in the LUCAS 3 group than in the ROSCER group during 5 minutes of CPR ( p < 0.001, p = 0.008, respectively). End-tidal CO 2 level of the ROSCER group was initially lower than that of the LUCAS 3 group, but was higher over time ( p = 0.022). A Kaplan-Meier survival analysis for ROSC also showed no difference between the two groups ( p = 0.46).

Conclusion

The prototype of a remote-controlled automated chest compression device can move the chest compression position without interruption during CPR. In a mannequin and a swine model of cardiac arrest, the device showed no inferior performance to a conventional mechanical CPR device.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cardiopulmonary ResuscitationHeart ArrestMaleAnimalsSwinePilot ProjectsManikinsPressureHemodynamics

Autor:innen und Institutionen

  • Department of Emergency Medicine, Seoul National University College of Medicine, Seoul, Republic of Korea.
  • Department of Emergency Medicine, Seoul National University Hospital, Seoul, Republic of Korea.
  • Research Center for Disaster Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea.
  • Graduate School of Convergence Science and Technology, Seoul National University, Seoul, Republic of Korea.
  • Advanced Institutes of Convergence Technology, Suwon, Republic of Korea.
  • NT Robot, Co, Seoul, Republic of Korea.

ORCID-Profile: Gil Joon Suh, Kyung Su Kim, Woon Yong Kwon, Jaehoon Sim, Jung Chan Lee, Woo Sang Cho

Förderung

  • Ministry of Health and Welfare

Zitieren

APA

Suh, G. J., Kim, T., Kim, K. S., Kwon, W. Y., Kim, H., Park, H., Wang, G., Park, J., Hur, S., Sim, J., Kim, K., Lee, J. C., Shin, D. A., Cho, W. S., Kim, B. J., Kwon, S., & Lee, Y. J. (2024). A remote-controlled automatic chest compression device capable of moving compression position during CPR: A pilot study in a mannequin and a swine model of cardiac arrest. PLOS ONE. https://doi.org/10.1371/journal.pone.0297057

Vancouver

Suh GJ, Kim T, Kim KS, Kwon WY, Kim H, Park H, et al. A remote-controlled automatic chest compression device capable of moving compression position during CPR: A pilot study in a mannequin and a swine model of cardiac arrest. PLOS ONE. 2024. doi:10.1371/journal.pone.0297057

BibTeX
@article{Suh2024,
  author = {Suh, Gil Joon and Kim, Taegyun and Kim, Kyung Su and Kwon, Woon Yong and Kim, Hayoung and Park, Heesu and Wang, Gaonsorae and Park, Jaeheung and Hur, Sungmoon and Sim, Jaehoon and Kim, Kyunghwan and Lee, Jung Chan and Shin, Dong Ah and Cho, Woo Sang and Kim, Byung Jun and Kwon, Soyoon and Lee, Ye Ji},