Intraocular lens simulator using computational holographic display for cataract patients
PLOS ONE · 2024
DOI: 10.1371/journal.pone.0295215 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS ONE, 2024. Verfasst von Deniz Akyazi und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Electronics Engineering, Koç University, Istanbul, Turkiye. Hauptthemen laut MEDLINE: Holography, Lenses, Intraocular, Cataract. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Türkiye Bilimsel ve Teknolojik Araştırma Kurumu, HORIZON EUROPE European Innovation Council.
Abstract (Originalsprache)
Purpose
To develop and validate a holography based vision simulator for the demonstration of expected postoperative vision corresponding to monofocal and multifocal intraocular lenses (IOL) to cataract patients before surgery.
Methods
An artificial eye model is used to measure the optical performance of different IOL types. The resultant aberrations and degradations are then modeled using phase holograms and shown to subjects on a holographic display. We measure the contrast and resolution loss, halos around the light sources, and point spread function (PSF) corresponding to three different IOLs. We tested the holography based vision simulator on 13 healthy subjects and 6 cataract patients.
Results
Monofocal, bifocal, and trifocal IOLs exhibited a contrast decrease of 5%, 42%, and 45% and a resolution limit of 4.49, 4.00, and 4.00 lp/mm (using 0.05 MTF criteria), respectively. Monofocal IOLs have the best resolution and contrast at the optimal focus distance, and multifocal lenses offer extended depth-of-field but exhibit prominent halos and reduced contrast/resolution.
Conclusion
We confirmed that the visual functions of IOLs could be successfully modeled using phase holograms and simulated using a holographic display without using a physical IOL. Patients can experience the effects of different IOL options prior to surgery, which helps with IOL selection, expectation management, and patient satisfaction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Electrical and Electronics Engineering, Koç University, Istanbul, Turkiye.
- School of Medicine, Koç University, Istanbul, Turkiye.
- Research Center for Translational Medicine, Koç University, Istanbul, Turkiye.
ORCID-Profile: Deniz Akyazi, Hakan Urey
Förderung
- Türkiye Bilimsel ve Teknolojik Araştırma Kurumu
- HORIZON EUROPE European Innovation Council
Zitieren
APA
Akyazi, D., Aygun, U., Sahin, A., & Urey, H. (2024). Intraocular lens simulator using computational holographic display for cataract patients. PLOS ONE. https://doi.org/10.1371/journal.pone.0295215
Vancouver
Akyazi D, Aygun U, Sahin A, Urey H. Intraocular lens simulator using computational holographic display for cataract patients. PLOS ONE. 2024. doi:10.1371/journal.pone.0295215
BibTeX
@article{Akyazi2024,
author = {Akyazi, Deniz and Aygun, Ugur and Sahin, Afsun and Urey, Hakan},
title = {Intraocular lens simulator using computational holographic display for cataract patients},
journal = {PLOS ONE},
year = {2024},
doi = {10.1371/journal.pone.0295215}
}
EUVIMED