MRI overestimates articular cartilage thickness and volume compared to synchrotron radiation phase-contrast imaging
PLOS ONE · 2023
DOI: 10.1371/journal.pone.0291757 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS ONE, 2023. Verfasst von Suranjan Bairagi und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anatomy, Physiology, and Pharmacology, University of Saskatchewan, Saskatoon, Saskatchewan, Canada. Hauptthemen laut MEDLINE: Cartilage, Articular, Osteoarthritis, Knee, Percutaneous Coronary Intervention. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: CIHR, royal university hospital foundation, Canadian Institutes of Health Research.
Abstract (Originalsprache)
Accurate evaluation of morphological changes in articular cartilage are necessary for early detection of osteoarthritis (OA). 3T magnetic resonance imaging (MRI) has highly sensitive contrast resolution and is widely used clinically to detect OA. However, synchrotron radiation phase-contrast imaging computed tomography (SR-PCI) can also provide contrast to tissue interfaces that do not have sufficient absorption differences, with the added benefit of very high spatial resolution. Here, MRI was compared with SR-PCI for quantitative evaluation of human articular cartilage. Medial tibial condyles were harvested from non-OA donors and from OA patients receiving knee replacement surgery. Both imaging methods revealed that average cartilage thickness and cartilage volume were significantly reduced in the OA group, compared to the non-OA group. When comparing modalities, the superior resolution of SR-PCI enabled more precise mapping of the cartilage surface relative to MRI. As a result, MRI showed significantly higher average cartilage thickness and cartilage volume, compared to SR-PCI. These data highlight the potential for high-resolution imaging of articular cartilage using SR-PCI as a solution for early OA diagnosis. Recognizing current limitations of using a synchrotron for clinical imaging, we discuss its nascent utility for preclinical models, particularly longitudinal studies of live animal models of OA.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Anatomy, Physiology, and Pharmacology, University of Saskatchewan, Saskatoon, Saskatchewan, Canada.
- Department of Surgery, University of Saskatchewan, Saskatoon, Saskatchewan, Canada.
- Department of Medical Imaging, University of Saskatchewan, Saskatoon, Saskatchewan, Canada.
- Canadian Light Source Inc., Saskatoon, Saskatchewan, Canada.
ORCID-Profile: Suranjan Bairagi, M Adam Webb, B Frank Eames
Förderung
- CIHR
- royal university hospital foundation
- Canadian Institutes of Health Research
- Saskatchewan Health Research Foundation
Zitieren
APA
Bairagi, S., Abdollahifar, M. A., Atake, O. J., Dust, W., Wiebe, S., Belev, G., Chapman, L. D., Webb, M. A., Zhu, N., Cooper, D. M. L., & Eames, B. F. (2023). MRI overestimates articular cartilage thickness and volume compared to synchrotron radiation phase-contrast imaging. PLOS ONE. https://doi.org/10.1371/journal.pone.0291757
Vancouver
Bairagi S, Abdollahifar MA, Atake OJ, Dust W, Wiebe S, Belev G, et al. MRI overestimates articular cartilage thickness and volume compared to synchrotron radiation phase-contrast imaging. PLOS ONE. 2023. doi:10.1371/journal.pone.0291757
BibTeX
@article{Bairagi2023,
author = {Bairagi, Suranjan and Abdollahifar, Mohammad-Amin and Atake, Oghenevwogaga J and Dust, William and Wiebe, Sheldon and Belev, George and Chapman, L Dean and Webb, M Adam and Zhu, Ning and Cooper, David M L and Eames, B Frank},
title = {MRI overestimates articular cartilage thickness and volume compared to synchrotron radiation phase-contrast imaging},
journal = {PLOS ONE},
year = {2023},
doi = {10.1371/journal.pone.0291757}
}