An approach to difficult airway in infants: Comparison of GlideScope® Spectrum LoPro, GlideScope® Spectrum Miller and conventional Macintosh and Miller blades in a simulated Pierre Robin sequence performed by 90 anesthesiologists
PLOS ONE · 2023
DOI: 10.1371/journal.pone.0288816 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS ONE, 2023. Verfasst von Andrea Irouschek und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology, Faculty of Medicine, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany. Hauptthemen laut MEDLINE: Pierre Robin Syndrome, Laryngoscopes, Soft Tissue Injuries. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft and Friedrich-Alexander-Universität Erlangen-Nürnberg.
Abstract (Originalsprache)
Background
Airway management can be challenging in neonates and infants. The Pierre Robin sequence (PRS) is a condition characterized by micrognathia, glossoptosis and airway obstruction. The airway management of these patients poses great challenges for anesthesiologists and pediatricians alike. To date, there has been no direct comparison of the hyperangulated GlideScope ® Spectrum LoPro (GLP), the straight GlideScope ® Spectrum Miller (GSM), a conventional Macintosh (MC) and a conventional Miller blade (ML) in patients with PRS.
Methods
For this purpose, 90 anesthesiologists (43 with limited experience, 47 with extensive experience) performed orotracheal intubation on an Air-Sim ® Pierre Robin X manikin using GLP, GSM, MC and ML in randomized order. ‘Time-to-vocal-cords’, ‘time-to-intubate’, ‘time-to-ventilate’, the severity of oral-soft-tissue-trauma and the subjective evaluation of each device were recorded.
Results
A significantly faster and better view of the vocal cords and lower oral-soft-tissue-trauma was achieved using the GLP (p120s) with the GLP. In the sub-group with extensive experience, we found significantly shorter intubation times for the GSM and ML. The GLP was the tool of choice for most participants, while the conventional MC received the lowest rating.
Conclusions
Videolaryngoscopy leads to increased safety for the prevention of undetected esophageal intubation in the airway management in a PRS manikin. Hyperangulated blades may ensure a good and fast view of the vocal cords and low oral-soft-tissue-trauma but pose a challenge during the placement of the tube. Specific skills and handling seem to be necessary to ensure a safe tube placement with this sort of blades.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Anesthesiology, Faculty of Medicine, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
ORCID-Profile: Andrea Irouschek
Förderung
- Deutsche Forschungsgemeinschaft and Friedrich-Alexander-Universität Erlangen-Nürnberg
Zitieren
APA
Irouschek, A., Moritz, A., Kremer, S., Fuchte, T., Danzl, A., Schmidt, J., & Golditz, T. (2023). An approach to difficult airway in infants: Comparison of GlideScope® Spectrum LoPro, GlideScope® Spectrum Miller and conventional Macintosh and Miller blades in a simulated Pierre Robin sequence performed by 90 anesthesiologists. PLOS ONE. https://doi.org/10.1371/journal.pone.0288816
Vancouver
Irouschek A, Moritz A, Kremer S, Fuchte T, Danzl A, Schmidt J, et al. An approach to difficult airway in infants: Comparison of GlideScope® Spectrum LoPro, GlideScope® Spectrum Miller and conventional Macintosh and Miller blades in a simulated Pierre Robin sequence performed by 90 anesthesiologists. PLOS ONE. 2023. doi:10.1371/journal.pone.0288816
BibTeX
@article{Irouschek2023,
author = {Irouschek, Andrea and Moritz, Andreas and Kremer, Sven and Fuchte, Tobias and Danzl, Anja and Schmidt, Joachim and Golditz, Tobias},
title = {An approach to difficult airway in infants: Comparison of GlideScope® Spectrum LoPro, GlideScope® Spectrum Miller and conventional Macintosh and Miller blades in a simulated Pierre Robin sequence performed by 90 anesthesiologists},
journal = {PLOS ONE},
year = {2023},
doi = {10.1371/journal.pone.0288816}
}
EUVIMED