EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Comparison of new and old BacT/ALERT aerobic bottles for detection of Candida species

Kye Won Choe iD, Yong Kwan Lim iD, Mi-Kyung Lee iD

PLOS ONE · 2023

DOI: 10.1371/journal.pone.0288674 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS ONE, 2023. Verfasst von Kye Won Choe und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, Chung-Ang University College of Medicine, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Candidemia, Bacteremia. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation of Korea (NRF) grant funded by the Korean government.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

A new version of aerobic blood culture media has been developed for the BacT/ALERT (bioMérieux) blood culture system. We evaluated the time to detection and yeast cell counts in positive blood cultures for each Candida spp. according to changes in media.

Methods

Isolates from defibrinated horse blood were inoculated into three types of bottles: the old version of aerobic bottle, new version of aerobic bottle, and anaerobic bottle. All bottles were incubated in the BacT/ALERT Virtuo blood culture system. The time to detection was monitored for each bottle, and yeast cell counts were performed immediately after testing positive, determined via the plate count method. Clinical retrospective data of the candidemia samples before and after aerobic bottle change also were analyzed.

Results

The median time to detection was 52.47 hours in the old aerobic bottles versus 19.92 hours in the new aerobic bottles ( P < 0.001) for Candida glabrata , and standard and clinical strains showed similar results. C. albicans (27.6 to 24.95 hours) and C. guilliermondii (28.92 to 26.9 hours) had shorter time to detection. However, C. auris (25.43 to 28.25 hours) had a longer time to detection in the new aerobic bottle. The retrospective clinical analysis showed a significant decrease in time to detection (45.0 to 19.4 hours) for C. glabrata , which is consistent with our simulated study result for C. glabrata . As a result of analysis including all blood specimens, C. tropicalis showed a significant delay in time to detection in new aerobic bottles. In an analysis limited to peripheral blood specimens, the time to detection of C. parapsilosis was longer in new aerobic bottles than in old aerobic bottles.

Conclusion

Most Candida species did not show remarkable TTD differences, but TTD of C. glabrata was markedly reduced in the New FA Plus bottle. The reduction of time to detection enables faster detection and therapeutic approach for C. glabrata infections.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

CandidemiaBacteremiaAnimalsHorsesCandidaRetrospective StudiesBlood CultureCandida albicansCulture MediaCandida glabrata

Substanzen

Culture Media

Autor:innen und Institutionen

  • Department of Laboratory Medicine, Chung-Ang University College of Medicine, Seoul, Republic of Korea.

ORCID-Profile: Kye Won Choe, Yong Kwan Lim, Mi-Kyung Lee

Förderung

  • National Research Foundation of Korea (NRF) grant funded by the Korean government

Zitieren

APA

Choe, K. W., Lim, Y. K., & Lee, M. K. (2023). Comparison of new and old BacT/ALERT aerobic bottles for detection of Candida species. PLOS ONE. https://doi.org/10.1371/journal.pone.0288674

Vancouver

Choe KW, Lim YK, Lee MK. Comparison of new and old BacT/ALERT aerobic bottles for detection of Candida species. PLOS ONE. 2023. doi:10.1371/journal.pone.0288674

BibTeX
@article{Choe2023,
  author = {Choe, Kye Won and Lim, Yong Kwan and Lee, Mi-Kyung},
  title = {Comparison of new and old BacT/ALERT aerobic bottles for detection of Candida species},
  journal = {PLOS ONE},
  year = {2023},
  doi = {10.1371/journal.pone.0288674}
}
Kostenlos starten →