Analysis of fixation materials in micro-CT: It doesn’t always have to be styrofoam

Jan Scherberich iD, Anton G Windfelder, Gabriele A Krombach

PLOS ONE · 14.06.2023 · 18 (6) · e0286039

DOI: 10.1371/journal.pone.0286039 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS ONE, online erschienen am 14.06.2023. Verfasst von Jan Scherberich und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology (Experimental Radiology), University Hospital Giessen, Giessen, Hesse, Germany. Hauptthemen laut MEDLINE: Polystyrenes, Polyethylene. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Uniklinikum Giessen und Marburg.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Good fixation of filigree specimens for micro-CT examinations is often a challenge. Movement artefacts, over-radiation or even crushing of the specimen can easily occur. Since different specimens have different requirements, we scanned, analysed and compared 19 possible fixation materials under the same conditions in the micro-CT. We focused on radiodensity, porosity and reversibility of these fixation materials. Furthermore, we have made sure that all materials are cheap and easily available. The scans were performed with a SkyScan 1173 micro-CT. All dry fixation materials tested were punched into 5 mm diameter cylinders and clamped into 0.2 ml reaction vessels. A voxel size of 5.33 μm was achieved in a 180° scan in 0.3° steps. Ideally, fixation materials should not be visible in the reconstructed image, i.e., barely binarised. Besides common micro-CT fixation materials such as styrofoam (-935 Hounsfield Units) or Basotect foam (-943 Hounsfield Units), polyethylene air cushions (-944 Hounsfield Units), Micropor foam (-926 Hounsfield Units) and polyurethane foam, (-960 Hounsfield Units to -470 Hounsfield Units) have proved to be attractive alternatives. Furthermore, more radiopaque materials such as paraffin wax granulate (-640 Hounsfield Units) and epoxy resin (-190 Hounsfield Units) are also suitable as fixation materials. These materials often can be removed in the reconstructed image by segmentation. Sample fixations in the studies of recent years are almost all limited to fixation in Parafilm, Styrofoam, or Basotect foam if the fixation type is mentioned at all. However, these are not always useful, as styrofoam, for example, dissolves in some common media such as methylsalicylate. We show that micro-CT laboratories should be equipped with various fixation materials to achieve high-level image quality.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PolystyrenesPolyethyleneX-Ray MicrotomographyArtifactsParaffin

Substanzen

styrofoamPolystyrenesPolyethyleneParaffin

Autor:innen und Institutionen

  • Department of Diagnostic and Interventional Radiology (Experimental Radiology), University Hospital Giessen, Giessen, Hesse, Germany.
  • Branch for Bioresources, Fraunhofer Institute for Molecular Biology and Applied Ecology IME, Giessen, Hesse, Germany.

ORCID-Profile: Jan Scherberich

Förderung

  • Uniklinikum Giessen und Marburg

Zitieren

APA

Scherberich, J., Windfelder, A. G., & Krombach, G. A. (2023). Analysis of fixation materials in micro-CT: It doesn’t always have to be styrofoam. PLOS ONE, 18(6), e0286039. https://doi.org/10.1371/journal.pone.0286039

Vancouver

Scherberich J, Windfelder AG, Krombach GA. Analysis of fixation materials in micro-CT: It doesn’t always have to be styrofoam. PLOS ONE. 2023;18(6):e0286039. doi:10.1371/journal.pone.0286039

BibTeX
@article{Scherberich2023,
  author = {Scherberich, Jan and Windfelder, Anton G and Krombach, Gabriele A},
  title = {Analysis of fixation materials in micro-CT: It doesn’t always have to be styrofoam},
  journal = {PLOS ONE},
  year = {2023},
  volume = {18},
  number = {6},
  pages = {e0286039},
  doi = {10.1371/journal.pone.0286039}
}
Kostenlos starten →