EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Test-retest reliability of the single leg stance on a Lafayette stability platform

Nureen Zaghlul iD, Siew Li Goh, Rizal Razman, Mahmoud Danaee iD, Chow Khuen Chan

PLOS ONE · 17.01.2023 · 18 (1) · e0280361

DOI: 10.1371/journal.pone.0280361 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS ONE, online erschienen am 17.01.2023. Verfasst von Nureen Zaghlul und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Faculty of Engineering, Universiti Malaya, Kuala Lumpur, Malaysia. Hauptthemen laut MEDLINE: Leg, Postural Balance. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 57 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universiti Malaya, Malaysia.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The validity and reliability of the Lafayette stability platform are well-established for double leg testing. However, no evaluation of single leg (SL) stance on the platform was discovered yet. Therefore, this study aimed to investigate the reliability of conducting the SL stance on the Lafayette platform. Thirty-six healthy and active university students (age 23.2 ± 3.2 years; BMI 21.1 ± 3.1 kg/m 2 ) were tested twice, one week apart (week 1; W1, week 2; W2). They stood on their dominant leg with eyes-open (EO) and eyes-closed (EC) in random order. Three successful trials of 20 seconds each were recorded. The duration during which the platform was maintained within 0° of tilt was referred to as time in balance (TIB). At all-time points, TIB was consistently longer in EO (EO W1 : 17.02 ± 1.04s; EO W2 : 17.32 ± 1.03s) compared to EC (EC W1 : 11.55 ± 1.73s; EC W2 : 13.08 ± 1.82s). A ±10 seconds difference was demonstrated in the Bland-Altman analysis in both EO and EC. Lower standard error of measurement (SEM) and coefficient of variation (CV) indicated consistent output. High intraclass correlation coefficient (ICC) values were seen between weeks (EO = 0.74; EC = 0.76) and within weeks (EO W1 = 0.79; EO W2 = 0.86; EC W1 = 0.71; EC W2 = 0.71). Although statistical measures (i.e., SEM, CV, and ICC) indicated good reliability of Lafayette for SL tasks, the wide agreement interval is yet to be clinically meaningful. Factors underlying the wide variation need to be identified before Lafayette is used for TIB assessment.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LegPostural BalanceHumansYoung AdultAdultReproducibility of ResultsEyePhysical Therapy Modalities

Autor:innen und Institutionen

  • Department of Biomedical Engineering, Faculty of Engineering, Universiti Malaya, Kuala Lumpur, Malaysia.
  • SEMREG, Sports Medicine Unit, Faculty of Medicine, Universiti Malaya, Kuala Lumpur, Malaysia.
  • Centre for Sport & Exercise Sciences, Universiti Malaya, Kuala Lumpur, Malaysia.
  • Department of Social and Preventive Medicine, Faculty of Medicine, Universiti Malaya, Kuala Lumpur, Malaysia.

ORCID-Profile: Nureen Zaghlul, Mahmoud Danaee

Förderung

  • Universiti Malaya, Malaysia

Zitieren

APA

Zaghlul, N., Goh, S. L., Razman, R., Danaee, M., & Chan, C. K. (2023). Test-retest reliability of the single leg stance on a Lafayette stability platform. PLOS ONE, 18(1), e0280361. https://doi.org/10.1371/journal.pone.0280361

Vancouver

Zaghlul N, Goh SL, Razman R, Danaee M, Chan CK. Test-retest reliability of the single leg stance on a Lafayette stability platform. PLOS ONE. 2023;18(1):e0280361. doi:10.1371/journal.pone.0280361

BibTeX
@article{Zaghlul2023,
  author = {Zaghlul, Nureen and Goh, Siew Li and Razman, Rizal and Danaee, Mahmoud and Chan, Chow Khuen},
  title = {Test-retest reliability of the single leg stance on a Lafayette stability platform},
  journal = {PLOS ONE},
  year = {2023},
  volume = {18},
  number = {1},
  pages = {e0280361},
  doi = {10.1371/journal.pone.0280361}
}
Kostenlos starten →