Fachartikel in der Zeitschrift PLOS Neglected Tropical Diseases, online erschienen am 23.02.2024. Verfasst von Brice Meulah und 13 weiteren Autor:innen. Beteiligte Einrichtung: Leiden University Center for Infectious Diseases (LUCID), Leiden University Medical Center, Leiden, The Netherlands. Hauptthemen laut MEDLINE: Artificial Intelligence, Microscopy, Schistosoma haematobium, Schistosomiasis haematobia. Die Arbeit wurde bisher 22-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NWO-WOTRO Science for Global Development program.
Abstract (Originalsprache)
Introduction
Schistosomiasis is a significant public health concern, especially in Sub-Saharan Africa. Conventional microscopy is the standard diagnostic method in resource-limited settings, but with limitations, such as the need for expert microscopists. An automated digital microscope with artificial intelligence (Schistoscope), offers a potential solution. This field study aimed to validate the diagnostic performance of the Schistoscope for detecting and quantifying Schistosoma haematobium eggs in urine compared to conventional microscopy and to a composite reference standard (CRS) consisting of real-time PCR and the up-converting particle (UCP) lateral flow (LF) test for the detection of schistosome circulating anodic antigen (CAA).
Methods
Based on a non-inferiority concept, the Schistoscope was evaluated in two parts: study A, consisting of 339 freshly collected urine samples and study B, consisting of 798 fresh urine samples that were also banked as slides for analysis with the Schistoscope. In both studies, the Schistoscope, conventional microscopy, real-time PCR and UCP-LF CAA were performed and samples with all the diagnostic test results were included in the analysis. All diagnostic procedures were performed in a laboratory located in a rural area of Gabon, endemic for S . haematobium .
Results
In study A and B, the Schistoscope demonstrated a sensitivity of 83.1% and 96.3% compared to conventional microscopy, and 62.9% and 78.0% compared to the CRS. The sensitivity of conventional microscopy in study A and B compared to the CRS was 61.9% and 75.2%, respectively, comparable to the Schistoscope. The specificity of the Schistoscope in study A (78.8%) was significantly lower than that of conventional microscopy (96.4%) based on the CRS but comparable in study B (90.9% and 98.0%, respectively).
Conclusion
Overall, the performance of the Schistoscope was non-inferior to conventional microscopy with a comparable sensitivity, although the specificity varied. The Schistoscope shows promising diagnostic accuracy, particularly for samples with moderate to higher infection intensities as well as for banked sample slides, highlighting the potential for retrospective analysis in resource-limited settings.
Trial registration
NCT04505046 ClinicalTrials.gov.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Leiden University Center for Infectious Diseases (LUCID), Leiden University Medical Center, Leiden, The Netherlands.
- Centre de Recherches Médicales des Lambaréné, CERMEL, Lambaréné, Gabon.
- Mechanical, Maritime and Material Engineering, Delft University of Technology, Delft, The Netherlands.
- Ecole doctorale régionale d'Afrique centrale en infectiologie tropicale de Franceville, Gabon.
- Fondation pour la Recherche Scientifique, Cotonou, Benin.
- Department of Cell and Chemical Biology, Leiden University Medical Center, Leiden, The Netherlands.
ORCID-Profile: Brice Meulah
Förderung
- NWO-WOTRO Science for Global Development program
Zitieren
APA
Meulah, B., Oyibo, P., Hoekstra, P. T., Moure, P. A. N., Maloum, M. N., Laclong-Lontchi, R. A., Honkpehedji, Y. J., Bengtson, M., Hokke, C., Corstjens, P. L. A. M., Agbana, T., Diehl, J. C., Adegnika, A. A., & van Lieshout, L. (2024). Validation of artificial intelligence-based digital microscopy for automated detection of Schistosoma haematobium eggs in urine in Gabon. PLOS Neglected Tropical Diseases, 18(2), e0011967. https://doi.org/10.1371/journal.pntd.0011967
Vancouver
Meulah B, Oyibo P, Hoekstra PT, Moure PAN, Maloum MN, Laclong-Lontchi RA, et al. Validation of artificial intelligence-based digital microscopy for automated detection of Schistosoma haematobium eggs in urine in Gabon. PLOS Neglected Tropical Diseases. 2024;18(2):e0011967. doi:10.1371/journal.pntd.0011967
BibTeX
@article{Meulah2024,
author = {Meulah, Brice and Oyibo, Prosper and Hoekstra, Pytsje T and Moure, Paul Alvyn Nguema and Maloum, Moustapha Nzamba and Laclong-Lontchi, Romeo Aime and Honkpehedji, Yabo Josiane and Bengtson, Michel and Hokke, Cornelis and Corstjens, Paul L A M and Agbana, Temitope and Diehl, Jan Carel and Adegnika, Ayola Akim and van Lieshout, Lisette},
title = {Validation of artificial intelligence-based digital microscopy for automated detection of Schistosoma haematobium eggs in urine in Gabon},
journal = {PLOS Neglected Tropical Diseases},
year = {2024},
volume = {18},
number = {2},
pages = {e0011967},
doi = {10.1371/journal.pntd.0011967}
}
EUVIMED