Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement

Alyssa Shapiro iD, Prince Mtenthaonga iD, Rowland Mjumira iD, Margaret Reuben, Ayodele Samuel, Meaghan Bond iD, Jennifer Carns iD, Richard Schwarz, Ryan Johnston, Lucky Mangwiro iD, Opeyemi Odedere, Robert Miros iD und 5 weitere

PLOS Global Public Health · 08.08.2023 · 3 (8) · e0002262

DOI: 10.1371/journal.pgph.0002262 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS Global Public Health, online erschienen am 08.08.2023. Verfasst von Alyssa Shapiro und 16 weiteren Autor:innen. Beteiligte Einrichtung: Department of Bioengineering, Rice University, Houston, Texas, United States of America. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: John D. and Catherine T. MacArthur Foundation, Bill and Melinda Gates Foundation, ELMA Foundation.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Neonatal jaundice is an important cause of morbidity and mortality worldwide, and neonates born in low and middle-income countries bear a disproportionate burden. We previously developed a low-cost, point-of-care system to measure total serum bilirubin (TSB) in neonates. This device was effective at detecting and monitoring jaundice; however, the disposable strips were difficult to produce at scale. Here, we report a new lateral flow cassette design, called BiliDx, that was produced at scale using traditional manufacturing techniques. We evaluated the performance of BiliDx at sites in Nigeria and Malawi. The lateral flow strip consists of plasma separation membranes, nitrocellulose, and a plastic cassette. We evaluated the performance of the strips and reader at two hospitals located in Nigeria and Malawi compared to reference standard TSB. We also assessed performance for samples with high direct bilirubin (DB) and high hematocrit (HCT). We collected 1,144 samples from 758 neonates (TSB ranged from 0.2 to 45.9 mg/dL). The mean bias of BiliDx measurements in the validation set was +0.75 mg/dL, and 95% limits of agreement were -2.57 to 4.07 mg/dL. The mean bias and limits of agreement were comparable for samples with HCT < 60% and HCT ≥ 60%, and for samples with low and intermediate DB levels; the samples with high DB levels had wider 95% limits of agreement (-4.50 to +3.03 mg/dL). Error grid analysis shows that 96.9% of samples measured with BiliDx would have resulted in the same clinical decision as the reference standard. This performance is comparable to previous results that used a handmade two-dimensional strip. Additionally, error grid analysis shows that all 20 samples with high DB levels would have resulted in the same clinical decision as the reference standard. This evaluation supports the use of BiliDx lateral flow cassettes to provide accurate point-of-care measurements in low-resource settings.

Autor:innen und Institutionen

  • Department of Bioengineering, Rice University, Houston, Texas, United States of America.
  • Rice360 Institute for Global Health Technologies, Rice University, Houston, Texas, United States of America.
  • Department of Pediatrics, Queen Elizabeth Central Hospital, Blantyre, Malawi.
  • Department of Paediatrics, College of Medicine, University of Lagos, Lagos, Nigeria.
  • Kamuzu University of Health Sciences, Blantyre, Malawi.
  • 3rd Stone Design, San Rafael, California, United States of America.

ORCID-Profile: Alyssa Shapiro, Prince Mtenthaonga, Rowland Mjumira, Meaghan Bond, Jennifer Carns, Lucky Mangwiro, Robert Miros, Rebecca Richards-Kortum

Förderung

  • John D. and Catherine T. MacArthur Foundation
  • Bill and Melinda Gates Foundation
  • ELMA Foundation
  • Children's Investment Fund Foundation
  • Lemelson Foundation
  • Ting Tsung and Wei Fong Chao Family Foundation
  • Sall Family Foundation
  • Individual donors to NEST 360

Zitieren

APA

Shapiro, A., Mtenthaonga, P., Mjumira, R., Reuben, M., Samuel, A., Bond, M., Carns, J., Schwarz, R., Johnston, R., Mangwiro, L., Odedere, O., Miros, R., McHugh, S., Kawaza, K., Dube, Q., Ezeaka, C., & Richards-Kortum, R. (2023). Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement. PLOS Global Public Health, 3(8), e0002262. https://doi.org/10.1371/journal.pgph.0002262

Vancouver

Shapiro A, Mtenthaonga P, Mjumira R, Reuben M, Samuel A, Bond M, et al. Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement. PLOS Global Public Health. 2023;3(8):e0002262. doi:10.1371/journal.pgph.0002262

BibTeX
@article{Shapiro2023,
  author = {Shapiro, Alyssa and Mtenthaonga, Prince and Mjumira, Rowland and Reuben, Margaret and Samuel, Ayodele and Bond, Meaghan and Carns, Jennifer and Schwarz, Richard and Johnston, Ryan and Mangwiro, Lucky and Odedere, Opeyemi and Miros, Robert and McHugh, Sean and Kawaza, Kondwani and Dube, Queen and Ezeaka, Chinyere and Richards-Kortum, Rebecca},
  title = {Design and field evaluation of a lateral flow cassette device for point-of-care bilirubin measurement},
  journal = {PLOS Global Public Health},
  year = {2023},
  volume = {3},
  number = {8},
  pages = {e0002262},
  doi = {10.1371/journal.pgph.0002262}
}
Kostenlos starten →