Data acquisition system for the measurement of the refractive index of air and the polarizability of CO 2 and N 2 using a Mach-Zehnder interferometer

Jorge A. Cibrián, Roberto E. Mejía

Optics Continuum · 15.11.2026 · 5 (11) · 3638

DOI: 10.1364/optcon.619018 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Optics Continuum, online erschienen am 02.11.2026. Verfasst von Jorge A. Cibrián und 1 weiteren Autor:innen. Die Arbeit nennt 15 Literaturangaben. Gefördert von: National Autonomous University of Honduras.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

We present a low-cost data acquisition system (DAS) based on the Arduino Due for high-precision gas refractometry. Integrated into a Mach-Zehnder interferometer, the DAS enables synchronous, 100 Hz sampling of optical irradiance and absolute pressure. We validated the system by measuring the molecular polarizability of CO 2 , N 2 , and air across four visible wavelengths in the range of 450–673 nm at T = 24 ∘ C, P = 80 kPa and 50% relative humidity. The experimental refractive index of air agrees with the corrected Edlén formulation. This open-source architecture bridges the gap between cost-prohibitive commercial systems and educational methods, spreading high-fidelity optical metrology tools.

Förderung

  • National Autonomous University of Honduras

Zitieren

APA

Cibrián, J. A., & Mejía, R. E. (2026). Data acquisition system for the measurement of the refractive index of air and the polarizability of CO 2 and N 2 using a Mach-Zehnder interferometer. Optics Continuum, 5(11), 3638. https://doi.org/10.1364/optcon.619018

Vancouver

Cibrián JA, Mejía RE. Data acquisition system for the measurement of the refractive index of air and the polarizability of CO 2 and N 2 using a Mach-Zehnder interferometer. Optics Continuum. 2026;5(11):3638. doi:10.1364/optcon.619018

BibTeX
@article{Cibrin2026,
  author = {Cibrián, Jorge A. and Mejía, Roberto E.},
  title = {Data acquisition system for the measurement of the refractive index of air and the polarizability of CO 2 and N 2 using a Mach-Zehnder interferometer},
  journal = {Optics Continuum},
  year = {2026},
  volume = {5},
  number = {11},
  pages = {3638},
  doi = {10.1364/optcon.619018}
}
Kostenlos starten →