This article reports the development of non-contact photoacoustic imaging (nC-PAI)—unlike the conventional PAI that mandates acoustic coupling gel—employing and integrating for the first time: (i) MEMS piezoelectric-micromachined-ultrasound-transducer (PMUT) sensor, (ii) Snell’s law-based refraction correction beamforming (S-rC-BForm), and (iii) signal averaging (sA). Quantitative pre-clinical validation studies—both numerical simulations and experiments—were conducted in diversified samples (agar phantoms and chicken breasts). The results demonstrate that our proposed nC-PAI enables achieving spatial-resolution ∼0.6mm, SNR ∼40dB, and contrast-ratio (CR) ∼45dB at depth ∼2.5mm (in tissues). Conclusively, the study demonstrates the promising practical applications of our fully contact- or coupling-free nC-PAI in clinical settings where one can not use the acoustic coupling gel, e.g., ophthalmic, skin-burns, intra-operative, wound-healing, and brain-imaging (non-intact skulls).
Förderung
- Department of Biotechnology
- Prime Minister’s Research Fellowship
Zitieren
APA
Paramanick, A., Kumar, A., Jeyaseelan, A., Samanta, D., Pratap, R., & Singh, M. S. (2026). Non-contact photoacoustic imaging empowered with a PMUT sensor and refraction correction. Optics Letters, 51(15), 4376. https://doi.org/10.1364/ol.603035
Vancouver
Paramanick A, Kumar A, Jeyaseelan A, Samanta D, Pratap R, Singh MS. Non-contact photoacoustic imaging empowered with a PMUT sensor and refraction correction. Optics Letters. 2026;51(15):4376. doi:10.1364/ol.603035
BibTeX
@article{Paramanick2026,
author = {Paramanick, Arijit and Kumar, Akshay and Jeyaseelan, Antony and Samanta, Deepayan and Pratap, Rudra and Singh, Mayanglambam Suheshkumar},
title = {Non-contact photoacoustic imaging empowered with a PMUT sensor and refraction correction},
journal = {Optics Letters},
year = {2026},
volume = {51},
number = {15},
pages = {4376},
doi = {10.1364/ol.603035}
}
EUVIMED