EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Calibration method for complex permittivity measurements using s-SNOM combining multiple probe tapping harmonics

Dario Siebenkotten, Bernd Kästner, Manuel Marschall, Arne Hoehl, Shuhei Amakawa

Optics Express · 17.06.2024 · 32 (13) · 23882

DOI: 10.1364/oe.523785 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Optics Express, online erschienen am 14.06.2024. Verfasst von Dario Siebenkotten und 4 weiteren Autor:innen. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministry of Education, Culture, Sports, Science and Technology, KAKENHI, Japan Society for the Promotion of Science.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Scattering-type scanning near-field optical microscopy (s-SNOM) enables sub-diffraction spectroscopy, featuring high sensitivity to small spatial permittivity variations of the sample surface. However, due to the complexity of the near-field probe-sample interaction, the quantitative extraction of the complex permittivity leads to a computationally demanding inverse problem, requiring further approximation of the system to an invertible model. Black-box calibration methods, similar to those applied to microwave vector network analyzers, allow the extraction of the permittivity without detailed electromagnetic modeling of the probe-sample interaction. These methods, however, are typically designed for stationary setups. In contrast, the distance between the sample and the probe tip of the s-SNOM is periodically modulated to differentiate the near-field interaction from the far-field background via lock-in detection of the harmonics of the periodic motion. This paper proposes an improved black-box calibration method that takes account of the effects of the probe tapping, including its multiple harmonics, and far-field background. The method is validated for an s-SNOM operating in the mid-infrared spectral range by applying it to spectroscopic measurements of silicon microstructures of different but well characterized doping.

Förderung

  • Ministry of Education, Culture, Sports, Science and Technology
  • KAKENHI
  • Japan Society for the Promotion of Science
  • Horizon 2020 Framework Programme

Zitieren

APA

Siebenkotten, D., Kästner, B., Marschall, M., Hoehl, A., & Amakawa, S. (2024). Calibration method for complex permittivity measurements using s-SNOM combining multiple probe tapping harmonics. Optics Express, 32(13), 23882. https://doi.org/10.1364/oe.523785

Vancouver

Siebenkotten D, Kästner B, Marschall M, Hoehl A, Amakawa S. Calibration method for complex permittivity measurements using s-SNOM combining multiple probe tapping harmonics. Optics Express. 2024;32(13):23882. doi:10.1364/oe.523785

BibTeX
@article{Siebenkotten2024,
  author = {Siebenkotten, Dario and Kästner, Bernd and Marschall, Manuel and Hoehl, Arne and Amakawa, Shuhei},
  title = {Calibration method for complex permittivity measurements using s-SNOM combining multiple probe tapping harmonics},
  journal = {Optics Express},
  year = {2024},
  volume = {32},
  number = {13},
  pages = {23882},
  doi = {10.1364/oe.523785}
}