Fachartikel in der Zeitschrift Biomedical Optics Express, online erschienen am 12.03.2024. Verfasst von Andres J Rodriguez und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Florida International University, Miami. Florida, 33174, USA. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: National Science Foundation.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Pulse oximetry represents a ubiquitous clinical application of optics in modern medicine. Recent studies have raised concerns regarding the potential impact of confounders, such as variable skin pigmentation and perfusion, on blood oxygen saturation measurement accuracy in pulse oximeters. Tissue-mimicking phantom testing offers a low-cost, well-controlled solution for characterizing device performance and studying potential error sources, which may thus reduce the need for costly in vivo trials. The purpose of this study was to develop realistic phantom-based test methods for pulse oximetry. Material optical and mechanical properties were reviewed, selected, and tuned for optimal biological relevance, e.g., oxygenated tissue absorption and scattering, strength, elasticity, hardness, and other parameters representing the human finger’s geometry and composition, such as blood vessel size and distribution, and perfusion. Relevant anatomical and physiological properties are summarized and implemented toward the creation of a preliminary finger phantom. To create a preliminary finger phantom, we synthesized a high-compliance silicone matrix with scatterers for embedding flexible tubing and investigated the addition of these scatterers to novel 3D printing resins for optical property control without altering mechanical stability, streamlining the production of phantoms with biologically relevant characteristics. Phantom utility was demonstrated by applying dynamic, pressure waveforms to produce tube volume change and resultant photoplethysmography (PPG) signals. 3D printed phantoms achieved more biologically relevant conditions compared to molded phantoms. These preliminary results indicate that the phantoms show strong potential to be developed into tools for evaluating pulse oximetry performance. Gaps, recommendations, and strategies are presented for continued phantom development.

Autor:innen und Institutionen

  • Department of Biomedical Engineering, Florida International University, Miami. Florida, 33174, USA.
  • Division of Biomedical Physics, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD 20993, USA.

ORCID-Profile: Andres J Rodriguez, William C Vogt, Christopher G Scully, T Joshua Pfefer

Förderung

  • National Science Foundation

Zitieren

APA

Rodriguez, A. J., Vasudevan, S., Farahmand, M., Weininger, S., Vogt, W. C., Scully, C. G., Ramella-Roman, J., & Pfefer, T. J. (2024). Tissue mimicking materials and finger phantom design for pulse oximetry. Biomedical Optics Express, 15(4), 2308. https://doi.org/10.1364/boe.518967

Vancouver

Rodriguez AJ, Vasudevan S, Farahmand M, Weininger S, Vogt WC, Scully CG, et al. Tissue mimicking materials and finger phantom design for pulse oximetry. Biomedical Optics Express. 2024;15(4):2308. doi:10.1364/boe.518967

BibTeX
@article{Rodriguez2024,
  author = {Rodriguez, Andres J and Vasudevan, Sandhya and Farahmand, Masoud and Weininger, Sandy and Vogt, William C and Scully, Christopher G and Ramella-Roman, Jessica and Pfefer, T Joshua},
  title = {Tissue mimicking materials and finger phantom design for pulse oximetry},
  journal = {Biomedical Optics Express},
  year = {2024},
  volume = {15},
  number = {4},
  pages = {2308},
  doi = {10.1364/boe.518967}
}
Kostenlos starten →