Visible-light optical coherence tomography platform for the characterization of the skin barrier
Biomedical Optics Express · 01.08.2023 · 14 (8) · 3914
DOI: 10.1364/boe.494356 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Optics Express, online erschienen am 06.07.2023. Verfasst von Dmitry G Revin und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electronic and Electrical Engineering, University of Sheffield, Sir Frederick Mappin Building, Sheffield, S1 3JD, UK. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Engineering and Physical Sciences Research Council.
Abstract (Originalsprache)
We demonstrate a free-space, trolley-mountable Fourier domain visible-light optical coherence tomography (OCT) system for studying the stratum corneum in non-palmar human skin. An axial resolution of 1 µm in tissue and at least −75 dB sensitivity have been achieved. High-quality B-scans, containing 1600 A-scans, are acquired at a rate of 39 Hz. Images from the dorsal hand, ventral wrist and ventral forearm areas are obtained, with a clearly resolved stratum corneum layer (typically 5–15 µm thick) presenting as a hypoechogenic dark layer below the bright entrance signal, similar to that found in palmar skin with traditional OCT systems. We find that the appearance of the stratum corneum layer strongly depends on its water content, becoming brighter after occlusive hydration.
Autor:innen und Institutionen
- Department of Electronic and Electrical Engineering, University of Sheffield, Sir Frederick Mappin Building, Sheffield, S1 3JD, UK.
- Dermatology Research, Department of Infection and Immunity and Cardiovascular Disease, University of Sheffield, Beech Hill Road, Sheffield, S10 2RX, UK.
- Currently with the Department of Radiology, the First Affiliated Hospital, Jinan University, No.613, Huangpu West Road, Tianhe District, Guangzhou, 510627, Guangdong, China.
ORCID-Profile: Robert A Byers
Förderung
- Engineering and Physical Sciences Research Council
Zitieren
APA
Revin, D. G., Byers, R. A., Duan, M. Q., Li, W., & Matcher, S. J. (2023). Visible-light optical coherence tomography platform for the characterization of the skin barrier. Biomedical Optics Express, 14(8), 3914. https://doi.org/10.1364/boe.494356
Vancouver
Revin DG, Byers RA, Duan MQ, Li W, Matcher SJ. Visible-light optical coherence tomography platform for the characterization of the skin barrier. Biomedical Optics Express. 2023;14(8):3914. doi:10.1364/boe.494356
BibTeX
@article{Revin2023,
author = {Revin, Dmitry G and Byers, Robert A and Duan, Meng Q and Li, Wei and Matcher, Stephen J},
title = {Visible-light optical coherence tomography platform for the characterization of the skin barrier},
journal = {Biomedical Optics Express},
year = {2023},
volume = {14},
number = {8},
pages = {3914},
doi = {10.1364/boe.494356}
}