Abstract (Originalsprache)

We investigated the feasibility of a mid-infrared laser heterodyne spectroscopy system in which the local oscillator (LO) laser light and incoherent light from natural sources are combined inside a coupler built from internally silver- and silver-iodide-coated silica capillary hollow fibers. Combining the two beams inside the fiber coupler removes the free-space beam splitter and the precise free-space alignment that conventional heterodyne spectrometers require, and therefore offers a simpler and mechanically more robust architecture for long-term ground-based monitoring and for compact instruments aboard spacecraft. With our test system, we showed the capability of the heterodyne spectrometer with the newly designed hollow fiber coupler for incoherent light. First, we succeeded in measuring an absorption line of C 2 H 4 at 10.3 µm in a gas cell with the incoherent background light from the blackbody with the ultrahigh resolution ( λ / d λ >10 6 ). The measured line profile agreed well with the model prediction. We also observed terrestrial atmospheric absorption of the transmitted sunlight. The measured intensity and spectral profile were consistent with the predicted spectra. The typical system noise temperature of the breadboard system was about 12,000 K, roughly four times higher than a conventional free-space beam-splitter configuration. A power-loss budget and a geometrical-optics analysis of the fiber output show that this gap is dominated by the throughput of the present coupler and that no penalty attributable to a loss of spatial coherence in the multimode hollow fiber is detected. These results demonstrated the applicability of the ultrahigh-resolution heterodyne spectrometer with the hollow fiber coupler to the observations of terrestrial and planetary atmospheres.

Förderung

  • Japan Society for the Promotion of Science

Zitieren

APA

Ohata, H., Nakagawa, H., Katagiri, T., Yamakawa, R., Kasaba, Y., Murata, I., Aida, H., Matsuura, Y., Hirahara, Y., & Yamazaki, A. (2026). Feasibility study of the ultrahigh-resolution mid-infrared laser heterodyne spectrometer with the hollow optical fiber coupler for incoherent lights. Applied Optics, 65(33), 11671. https://doi.org/10.1364/ao.610067

Vancouver

Ohata H, Nakagawa H, Katagiri T, Yamakawa R, Kasaba Y, Murata I, et al. Feasibility study of the ultrahigh-resolution mid-infrared laser heterodyne spectrometer with the hollow optical fiber coupler for incoherent lights. Applied Optics. 2026;65(33):11671. doi:10.1364/ao.610067

BibTeX
@article{Ohata2026,
  author = {Ohata, Hajime and Nakagawa, Hiromu and Katagiri, Takashi and Yamakawa, Ryo and Kasaba, Yasumasa and Murata, Isao and Aida, Haruki and Matsuura, Yuji and Hirahara, Yasuhiro and Yamazaki, Atsushi},
  title = {Feasibility study of the ultrahigh-resolution mid-infrared laser heterodyne spectrometer with the hollow optical fiber coupler for incoherent lights},
  journal = {Applied Optics},
  year = {2026},
  volume = {65},
  number = {33},
  pages = {11671},
  doi = {10.1364/ao.610067}
}
Kostenlos starten →