EUVIMEDEUVIMED entdecken
Open Access

Reducing False Alarm Rates and Workload in ICUs by Improving Arrhythmia Detection Algorithms of Patient Monitoring Systems

Jiyoon Oh iD, You Rim Kim, Yong Ju Lee, Dae Keun Jeon, Ji Soo Choi iD, Jae-Sun Uhm iD, Chul-Min Ahn iD, Won Seuk Jang iD

Medical Science Monitor · 05.12.2025 · 31

DOI: 10.12659/msm.949932 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Science Monitor, online erschienen am 05.12.2025. Verfasst von Jiyoon Oh und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Device Engineering and Management, Yonsei University College of Medicine, Seoul, South Korea. Hauptthemen laut MEDLINE: Clinical Alarms, Arrhythmias, Cardiac. Die Arbeit wurde bisher 2-mal zitiert (Crossref). Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND Patient monitoring systems are widely used in intensive care units (ICUs) to monitor patient's conditions. A high false alarm rate can lead to alarm fatigue among nurses, increasing workload and stress. This study aimed to improve the accuracy of arrhythmia detection by enhancing the noise detection algorithm in patient monitoring systems and to determine whether false alarm rate and workload decreased through clinical trials. MATERIAL AND METHODS Trials were conducted on adult patients in the ICU at Yongin Severance Hospital who required continuous electrocardiogram (ECG) monitoring for at least 2 days. After the first trial, the noise detection algorithm of the M50 (investigational device) was improved, and a second trial was conducted to evaluate its performance. Both trials followed the same study design. During the study period, M50 and MX700 (comparator device) were applied simultaneously for 3 days. Arrhythmia alarms were reviewed by an independent evaluator who assessed false alarms by comparing them with the ECG signals. False alarm rates were compared between trials using the chi-square (χ²) test. RESULTS The clinical trial was conducted through 2 separate trials, with 17 and 11 participants, respectively. A comparative analysis of false alarm rates of the investigational device demonstrated a reduction from 71.75% to 27.61%. Statistical analysis using the chi-square test indicated a P value of 0.000 (

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Clinical AlarmsArrhythmias, CardiacHumansAlgorithmsIntensive Care UnitsMonitoring, PhysiologicElectrocardiographyWorkloadAdultFemaleMaleMiddle AgedAgedDetection Algorithms

Autor:innen und Institutionen

  • Department of Medical Device Engineering and Management, Yonsei University College of Medicine, Seoul, South Korea.
  • Medical Device Usability Research Center, Gangnam Severance Hospital, Yonsei University College of Medicine, Seoul, South Korea.
  • Patient monitoring Lab, Mediana Co., Ltd, Seoul, South Korea.
  • Division of Pulmonology, Allergy and Critical Care Medicine, Department of Internal Medicine, Yongin Severance Hospital, Yonsei University College of Medicine, Yongin, South Korea.
  • Department of Cardiology, Severance Hospital, Yonsei University College of Medicine, Seoul, South Korea.
  • Division of Cardiology, Department of Internal Medicine, Severance Cardiovascular Hospital, Yonsei University Health System, Seoul, South Korea.

ORCID-Profile: Jiyoon Oh, Ji Soo Choi, Jae-Sun Uhm, Chul-Min Ahn, Won Seuk Jang

Zitieren

APA

Oh, J., Kim, Y. R., Lee, Y. J., Jeon, D. K., Choi, J. S., Uhm, J. S., Ahn, C. M., & Jang, W. S. (2025). Reducing False Alarm Rates and Workload in ICUs by Improving Arrhythmia Detection Algorithms of Patient Monitoring Systems. Medical Science Monitor, 31. https://doi.org/10.12659/msm.949932

Vancouver

Oh J, Kim YR, Lee YJ, Jeon DK, Choi JS, Uhm JS, et al. Reducing False Alarm Rates and Workload in ICUs by Improving Arrhythmia Detection Algorithms of Patient Monitoring Systems. Medical Science Monitor. 2025;31. doi:10.12659/msm.949932

BibTeX
@article{Oh2025,
  author = {Oh, Jiyoon and Kim, You Rim and Lee, Yong Ju and Jeon, Dae Keun and Choi, Ji Soo and Uhm, Jae-Sun and Ahn, Chul-Min and Jang, Won Seuk},
  title = {Reducing False Alarm Rates and Workload in ICUs by Improving Arrhythmia Detection Algorithms of Patient Monitoring Systems},
  journal = {Medical Science Monitor},
  year = {2025},
  volume = {31},
  doi = {10.12659/msm.949932}
}