A novel technique to quantify bone-to-implant contact of zygomatic implants: a radiographic analysis based on three-dimensional image registration and segmentation
Dentomaxillofacial Radiology · 2023
DOI: 10.1259/dmfr.20220210 ↗Kurz gefasst
Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, 2023. Verfasst von Yuwei Gu und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Second Dental Center and Oral Rehabilitation, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine; Colle…. Hauptthemen laut MEDLINE: Dental Implants. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Objectives
The purpose of this study is to establish a novel, reproducible technique to obtain the BIC area (BICA) between zygomatic implants and zygomatic bone based on post-operative cone-beam computed tomography (CBCT) images. Three-dimensional (3D) image registration and segmentation were used to eliminate the effect of metal-induced artifacts of zygomatic implants.
Methods
An ex-vivo study was included to verify the feasibility of the new method. Then, the radiographic bone-to-implant contact (rBIC) of 143 implants was measured in a total of 50 patients. To obtain the BICA of zygomatic implants and the zygomatic bone, several steps were necessary, including image preprocessing of CBCT scans, identification of the position of zygomatic implants, registration, and segmentation of pre- and post-operative CBCT images, and 3D reconstruction of models. The conventional two-dimensional (2D) linear rBIC (rBIC c ) measurement method with post-operative CBCT images was chosen as a comparison.
Results
The mean values of rBIC and rBIC c were 15.08 ± 5.92 mm and 14.77 ± 5.14 mm, respectively. A statistically significant correlation was observed between rBIC and rBIC c values ([Formula: see text]=0.86, p < 0.0001).
Conclusions
This study proposed a standardized, repeatable, noninvasive technique to quantify the rBIC of post-operative zygomatic implants in 3D terms. This technique is comparable to conventional 2D linear measurements and seems to be more reliable than these conventional measurements; thus, this method could serve as a valuable tool in the performance of clinical research protocols.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Second Dental Center and Oral Rehabilitation, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine; College of Stomatology, Shanghai Jiao Tong University; National Center for…
- Department of Stomatology, Shanghai Tenth People's Hospital, Tongji University School of Medicine, Shanghai, PR China.
- Institute of Biomedical Manufacturing and Life Quality Engineering, State Key Laboratory of Mechanical System and Vibration, School of Mechanical Engineering, Shanghai Jiao Tong University, Shanghai, China.
- Department of Oral Implantology, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine; College of Stomatology, Shanghai Jiao Tong University; National Center for Stomatology; National Clini…
- Institute of Medical Robotics, Shanghai Jiao Tong University, Shanghai, China.
Zitieren
APA
Gu, Y., Zhang, D., Tao, B., Wang, F., Chen, X., & Wu, Y. (2023). A novel technique to quantify bone-to-implant contact of zygomatic implants: a radiographic analysis based on three-dimensional image registration and segmentation. Dentomaxillofacial Radiology. https://doi.org/10.1259/dmfr.20220210
Vancouver
Gu Y, Zhang D, Tao B, Wang F, Chen X, Wu Y. A novel technique to quantify bone-to-implant contact of zygomatic implants: a radiographic analysis based on three-dimensional image registration and segmentation. Dentomaxillofacial Radiology. 2023. doi:10.1259/dmfr.20220210
BibTeX
@article{Gu2023,
author = {Gu, Yuwei and Zhang, Dingzhong and Tao, Baoxin and Wang, Feng and Chen, Xiaojun and Wu, Yiqun},
title = {A novel technique to quantify bone-to-implant contact of zygomatic implants: a radiographic analysis based on three-dimensional image registration and segmentation},
journal = {Dentomaxillofacial Radiology},
year = {2023},
doi = {10.1259/dmfr.20220210}
}
EUVIMED