Geometric validation of a computer simulator used in radiography education
BJR|Open · 2020-11 · 2 (1) · 20190027
DOI: 10.1259/bjro.20190027 ↗Kurz gefasst
Fachartikel in der Zeitschrift BJR|Open, 2020-11. Verfasst von Philip Cosson und 1 weiteren Autor:innen. Beteiligte Einrichtung: Teesside University, Middlesbrough, United Kingdom. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Objectives
The radiographical process of projection of a complex human form onto a two-dimensional image plane gives rise to distortions and magnifications. It is important that any simulation used for educational purposes should correctly reproduce these. Images generated using a commercially available computer simulation widely used in radiography education (ProjectionVR TM ) were tested for geometric accuracy of projection in all planes.
Methods
An anthropomorphic skull phantom was imaged using standard projection radiography techniques and also scanned using axial CT acquisition. The data from the CT was then loaded into the simulator and the same projection radiography techniques simulated. Bony points were identified on both the real radiographs and the digitally reconstructed radiographs (DRRs). Measurements sensitive to rotation and magnification were chosen to check for rotation and distortion errors.
Results
The real radiographs and the DRRs were compared by four experienced observers and measurements taken between the identified bony points on each of the images obtained. Analysis of the mean observations shows that the measurement from the DRR matches the real radiograph +1.5 mm/−1.5 mm. The Bland Altman bias was 0.55 (1.26 STD), with 95% limits of agreement 3.01 to −1.91.
Conclusions
Agreement between the empirical measurements is within the reported error of cephalometric analysis in all three anatomical planes. The image appearances of both the real radiographs and DRRs compared favourably. Advances in knowledge: The commercial computer simulator under test (ProjectionVR TM ) was able to faithfully recreate the image appearances of real radiography techniques, including magnification and distortion. Students using this simulation for training will obtain feedback likely to be useful when lessons are applied to real-world situations.
Autor:innen und Institutionen
- Teesside University, Middlesbrough, United Kingdom.
ORCID-Profile: Philip Cosson
Zitieren
APA
Cosson, P., & Lu, Z. (2020). Geometric validation of a computer simulator used in radiography education. BJR|Open, 2(1), 20190027. https://doi.org/10.1259/bjro.20190027
Vancouver
Cosson P, Lu Z. Geometric validation of a computer simulator used in radiography education. BJR|Open. 2020;2(1):20190027. doi:10.1259/bjro.20190027
BibTeX
@article{Cosson2020,
author = {Cosson, Philip and Lu, Zenghai},
title = {Geometric validation of a computer simulator used in radiography education},
journal = {BJR|Open},
year = {2020},
volume = {2},
number = {1},
pages = {20190027},
doi = {10.1259/bjro.20190027}
}
EUVIMED